Nikon DS-L4 Instructions Manual page 110

Microscope camera control unit
Hide thumbs Also See for DS-L4:
Table of Contents

Advertisement

Chapter 10
To use calibration settings of the DS-L4 (instead of
the microscope's objective information)
If necessary, you can use calibration settings that are
locally registered in the DS-L4, instead of the objective
information registered in the motorized microscope. For
example, you may want to attach infrequently used
objectives to the microscope temporarily. In such a
case, by replacing the DS-L4 local calibration data with
the information about those objectives, the calibration
settings can be automatically changed by simply
changing the nosepiece. (The calibration settings must
have been registered before use. For details on how to
register them, refer to "10.1.3 (2) Registering local
calibration settings".)
(1) Display the [MEAS/DRAW] menu. Tap the desired
measurement or drawing button.
(2) Tap [CALIB].
(3) Tap [OBJ LIST REPLACE].
(4) Select the address of the objective whose calibration setting you want to replace, and the number of the local
calibration setting registered in the DS-L4. Then, tap [REPLACE]. The calibration data is replaced.
To reset the calibration data to the microscope's objective data, select the address of the relevant objective,
and then tap [MIC CALIB].
To reset the calibration data of all objectives, tap [ALL MIC CALIB].
(5) Tap [SAVE] to save the new settings.
Current assignment
of objective
addresses
When you turn on the microscope after replacing the calibration data, a message "The calibration data has been
changed. Reset to the MIC CALIB?" appears on the DS-L4. To use the microscope's objective data, select [Yes].
To use the local calibration settings saved in the DS-L4, select [No].
Preparation for Using the Annotation Function and Measurement Function
Calibration data registered
in the microscope
102
Tap this
button to
replace a
calibration
setting.
Select an objective
address and local
calibration setting
number, and then
tap [REPLACE].
Calibration data locally
saved in the DS-L4

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents