7.. Horizontal machine alignment (Change 1)

On pages 50-51, in the Mount wireless module, sensor and prism section,
replace Step 1 through Step 4 including Figure 7-2, then continue the procedure
with Steps 5:
Mount the Wireless Sensor and prism
1. Locate the bracket fixed on the shaft of the left machine (usually stationary
machine) – as viewed from normal working position.
2. Mount the Sensor on the support posts of this bracket.
Note the orientation of the Sensor. The micro USB port must be at the
top. (See
Loosen the yellow knobs until you can slide the housing onto the support
posts.
3. Tighten the yellow knobs to clamp the Sensor onto the support posts. Do not
overtighten the yellow locking knobs. (See
Item
B.
Make sure that the laser beam can pass over or through the coupling and
is not blocked.