Agilent Technologies G1330B Reference Manual page 69

1100 series autosampler thermostat
Table of Contents

Advertisement

Repairing the ALS Thermostat Module
Exchanging the Sensors
7 Reinsert the sealing adapter by carefully
sliding it in to the seat.
9 Route the cables in the guide channels and
reconnect the connectors in the correct order
(See label on Power Supply).
8 Reinstall the hardfoam encapsulation.
10 Reinstall the Top Plate and the Top Cover.
69