Lost Contact To Als Thermostat - Agilent Technologies G1330B Reference Manual

1100 series autosampler thermostat
Table of Contents

Advertisement

Probable Causes:
Suggested Actions:
Troubleshooting and Test Functions

Lost Contact to ALS Thermostat

Lost Contact to ALS Thermostat
The autosampler control electronics continuously checks whether the ALS
thermostat is active or not.
• Bad cable connection between autosampler and ALS thermostat
• Cable between autosampler and thermostat module not connected
• Defective electronic board in autosampler or ALS thermostat
❏ Check cable connection between autosampler and ALS thermostat
❏ Connect cable between autosampler and ALS thermostat
❏ Exchange ASM board in the autosampler
❏ Exchange TCA board in the thermostat
❏ contact the Agilent Technologies service organization