Agilent Technologies G1330B Reference Manual

Agilent Technologies G1330B Reference Manual

1100 series autosampler thermostat
Table of Contents

Advertisement

Agilent 1100 Series
Autosampler Thermostat
Reference Manual
s2

Advertisement

Table of Contents
loading

Summary of Contents for Agilent Technologies G1330B

  • Page 1 Agilent 1100 Series Autosampler Thermostat Reference Manual...
  • Page 2 Edition 08/01 manual in order to protect the apparatus against damage. Agilent Technologies shall not Printed in Germany be liable for errors contained herein or for incidental or consequential damages in...
  • Page 3 Agilent 1100 Series Autosampler Thermostat Reference Manual...
  • Page 4 In This Book This manual contains information about the Agilent 1100 Series thermostatted autosamplers. Information about the autosamplers can be found in the respective Reference Manuals. This supplemental manual describes only the parts of the autosamplers that are different from the standard versions.
  • Page 5: Table Of Contents

    ALS Thermostat Operation 12 ALS Thermostat Control and Electronics 14 Temperature Controlled Autosampler Board (TCA) 15 Electrical Connections 17 Installing the G1330B Thermostat 19 Site requirements and installation of the G1330B thermostat 19 Site Requirements 20 Power Consideration 20 Power Cords 20...
  • Page 6 Contents Stage 3: Flow Connections 33 Stage 4: Installing the Sample Tray 34 Stage 5: Installing Tray Cover and Front Cover 36 Stage 6: Turning on the Thermostatted Autosampler 37 Stage 7: Update of Control Module Firmware 37 Stage 8: Update of the Agilent ChemStation Software 38 Transporting the Thermostatted Autosampler 40 Troubleshooting and Test Functions 41 The thermostatted autosampler‘s built-in troubleshooting...
  • Page 7 Contents Repairing the ALS Thermostat Module 55 Instructions on repair procedures of the ALS thermostat 55 Simple Repairs 56 Exchanging Internal Parts 56 Cleaning the Thermostat 57 Using the ESD Strap 58 Using the ESD Strap 58 Overview 59 Exchanging the Power Supply Fuses 60 Removing the Top Cover and Foam 61 Exchanging Heat Exchanger Fan 63 Exchanging the Heatsink Fans 65...
  • Page 8 Contents Specifications 83 Performance specifications of the ALS thermostat 83 Performance Specifications 84 Safety Information 90 Radio Interference 93 Sound Emission 93 Agilent Technologies on Internet 94...
  • Page 9: Theory Of Operation

    Theory of Operation Theory of operation of mechanical hardware, electronics, and instrument interfaces...
  • Page 10: Introduction To The Thermostatted Autosampler

    Theory of Operation Introduction to the Thermostatted Autosampler Introduction to the Thermostatted Autosampler The Agilent 1100 Series thermostatted autosampler is designed for use with other modules of the Agilent 1100 Series LC system, with the HP 1050 Series, or with other LC systems if adequate remote control inputs and outputs are available.
  • Page 11 Theory of Operation Introduction to the Thermostatted Autosampler Figure 1 Overview of the ALS Thermostat Power Supply TCA Board Heat sink fans Heat sink Heat exchanger Condensation leak drain Heat exchanger Fan...
  • Page 12: Als Thermostat Operation

    Theory of Operation ALS Thermostat Operation ALS Thermostat Operation Figure 2 ALS Thermostat Principle heat exchanger special sample tray air channel in autosampler sampling unit air outlet Peltier elements heating/cooling module ambient air in thermostatted air heated air out The thermostatted autosampler is equipped with a cooling/heating module which uses Peltier elements for efficient air cooling.
  • Page 13 Theory of Operation ALS Thermostat Operation underneath the special sample tray where it is evenly distributed and streams back into the sample tray area. From there it is again drawn into the ALS thermostat. This “recycle” mode assures a very efficient cooling/heating of the sample vials.
  • Page 14: Als Thermostat Control And Electronics

    Theory of Operation ALS Thermostat Control and Electronics ALS Thermostat Control and Electronics The ASM board of the autosampler module provides all control signals for the ALS thermostat (e.g., temperature setting) and receives the sensor signals from the temperature and fan sensors. The TCA board controls the internal assemblies in the ALS thermostat.
  • Page 15: Temperature Controlled Autosampler Board (Tca)

    Theory of Operation Temperature Controlled Autosampler Board (TCA) Temperature Controlled Autosampler Board (TCA) The board controls all information and activities of all assemblies within the ALS thermostat. The operator enters parameters, changes modes and controls the module through the control module or Agilent ChemStation, that is connected to the autosampler module.
  • Page 16 Theory of Operation Temperature Controlled Autosampler Board (TCA) multiplexer. The output of the multiplexer is connected to the autosampler main board via the 26 pin autosampler to ALS thermostat cable. The autosampler firmware checks the signals for correctness and initiates all needed activities.
  • Page 17: Electrical Connections

    Theory of Operation Electrical Connections Electrical Connections WAR NI N G Never use cables other than the ones supplied by Agilent Technologies to ensure proper functionality and compliance with safety or EMC regulations. Figure 5 Elctrical Connections CAN cable to next module...
  • Page 18 • The REMOTE connector may be used in combination with other analytical instruments from Agilent Technologies if you want to use features such as common shut down, prepare, and so on. • The RS-232 connector may be used to control the thermostatted autosampler from a computer through an RS-232 connection, using appropriate software.
  • Page 19: Installing The G1330B Thermostat

    Installing the G1330B Thermostat Site requirements and installation of the G1330B thermostat...
  • Page 20: Site Requirements

    The thermostatted autosamplers comprises two modules, the autosampler module (G1329A, G1389A, G1367A, or G2260A) and the thermostat module (G1330B Therm). Both modules have a separate power supply and a power plug for the line connections. The two modules are connected by a control cable and both are turned on by the autosampler module.
  • Page 21: Bench Space

    Installing the G1330B Thermostat Site Requirements on the power cords which connect to the rear of the two instrument is identical for all types of power cord. WAR NI N G Never operate your instrumentation from a power outlet that has no ground connection.
  • Page 22 Installing the G1330B Thermostat Site Requirements Table 1 Physical Specifications - Thermostatted Autosampler Type Specification Comments Thermostat Weight 20.7 kg (45.6 lbs) Dimensions 140 × 345 × 435 mm (height × width × depth) (5.5 × 13.5 × 17 inches) Line voltage ±...
  • Page 23: Unpacking The Autosampler

    Installing the G1330B Thermostat Unpacking the Autosampler Unpacking the Autosampler CA UT IO N If you intend to ship or transport the thermostatted autosampler, always park the transport assembly of the autosampler module before shipment (see “Transporting the Autosampler” in the corresponding manual).
  • Page 24: Delivery Checklist

    ALS thermostat. The delivery checklist are shown in Table 2. Please report missing or damaged parts to your local Agilent Technologies sales and service office. If the thermostatted autosampler was ordered as an upgrade (G1395A) to an existing autosampler, the shipment will also contain the required software upgrades for your Agilent ChemStation.
  • Page 25: Optimizing The Stack Configuration

    Installing the G1330B Thermostat Optimizing the Stack Configuration Optimizing the Stack Configuration If your thermostatted autosampler is part of a system, you can ensure optimum performance by installing the thermostatted autosampler in the stack in the position shown in Figure 6 and Figure 7. This configuration optimizes the system flow path, ensuring minimum delay volume.
  • Page 26 Installing the G1330B Thermostat Optimizing the Stack Configuration Figure 7 Recommended Stack Configuration (Rear View) Analog signal to recorder CAN bus cable Autosampler - Thermostat cable Remote cable Analog signal AC power AC power to recorder GPIB to LC ChemStation...
  • Page 27: Controller Requirements

    PCMCIA card that has the newest revision loaded. The firmware is not part of the shipment of the thermostatted autosampler. Contact your local Agilent Technologies sales and service office for the firmware update of the control module. For the update procedure see “Stage 7: Update of Control Module Firmware”...
  • Page 28: Installing The G1330B Thermostat

    Installing the G1330B Thermostat Installing the G1330B Thermostat Installing the G1330B Thermostat WAR NI N G To avoid personal injury, keep fingers away from the needle area during thermostatted autosampler operation. Do not bend the safety flap away from its position, or attempt to remove the safety cover (see Figure 8).
  • Page 29 Installing the G1330B Thermostat Installing the G1330B Thermostat Figure 9 Preparation of the ALS Thermostat Module locks Waste tube assembly Condensation leak tube 4 Connect the condensation leak tube to the main waste exit of the ALS thermostat and place into an appropriate vessel. It is possible to either let the condensation leak tubing exit the module at the front or at the left side of the module.
  • Page 30 Installing the G1330B Thermostat Installing the G1330B Thermostat Figure 10 Condensation Leak outlet Condensation leak tube Waste bottle 5 Install the front cover of the ALS thermostat. 6 Place the autosampler module on top of the ALS thermostat. Make sure that the autosampler is correctly engaged in the ALS thermostat locks.
  • Page 31: Stage 2: Power Cable And Interface Cable Connection

    Installing the G1330B Thermostat Installing the G1330B Thermostat Figure 11 Preparation of ALS Thermostat and Autosampler Air channel adapter Stage 2: Power Cable and Interface Cable Connection 1 Ensure the power switch on the front of the autosampler is OFF and the power cables are disconnected.
  • Page 32 Installing the G1330B Thermostat Installing the G1330B Thermostat 6 If required, connect additional interface and control cables to the thermostatted autosampler (see Figure 7 and Figure 13). Refer to the documentation of the Agilent 1100 Series control module or ChemStation for LC for more information.
  • Page 33: Stage 3: Flow Connections

    Installing the G1330B Thermostat Installing the G1330B Thermostat Figure 13 Cable Connections. Relay contacts CAN cable to next module Vial number output Therm-ALS cable Remote RS232C CAN cable to previous module CAN-bus GPIB Thermostat - Autosampler cable Stage 3: Flow Connections WAR NI N G When opening capillary or tube fittings, solvents may leak out.
  • Page 34: Stage 4: Installing The Sample Tray

    Installing the G1330B Thermostat Installing the G1330B Thermostat 1 Connect the pump outlet capillary to port 1 of the injection valve. 2 Connect column-compartment inlet capillary to port 6 of the injection valve. 3 Ensure that the waste tube is positioned inside the leak channel.
  • Page 35 Installing the G1330B Thermostat Installing the G1330B Thermostat Figure 15 Installing the Sample Tray Half-Tray Combinations N O T E In the thermostatted autosampler only the 100 vial tray is supported for temperature control of the vials. Nevertheless the half trays of the standard autosampler (G1313A) can be used in the thermostatted autosampler as well.
  • Page 36: Stage 5: Installing Tray Cover And Front Cover

    Installing the G1330B Thermostat Installing the G1330B Thermostat Numbering of Vial Positions The standard 100-vial tray has vial positions 1 to 100. However, when using two half-trays, the numbering convention is slightly different. The vial positions of the right-hand half tray begin at position 101 as follows: Left-hand 40-position tray: 1–40...
  • Page 37: Stage 6: Turning On The Thermostatted Autosampler

    Installing the G1330B Thermostat Installing the G1330B Thermostat Figure 17 Installation of Tray Cover and Front Cover of the Autosampler Tray cover Autosampler front door Stage 6: Turning on the Thermostatted Autosampler 1 Depress the power switch to turn on the two modules of the thermostatted autosampler.
  • Page 38: Stage 8: Update Of The Agilent Chemstation Software

    Installing the G1330B Thermostat Installing the G1330B Thermostat If you have a control module version A.01.30 or lower, update the firmware as described. 1 Disconnect the control module before inserting the PC card. 2 Insert the PC card into the card slot of the control module.
  • Page 39 Installing the G1330B Thermostat can be obtained from the Internet (http://www.chem.agilent.com/cag/ literature/apglit.html) or your local Agilent Technologies sales office. 1 If you decide to update your A.04.01 Agilent ChemStation to A.05.01, use the provided A.05.01 CD-ROM and follow the steps described in the Installing your ChemStation manual, provided as portable document format (PDF) file on the A.05.01 CD-ROM in the directory MANUALS\INSTALL\LC.
  • Page 40: Transporting The Thermostatted Autosampler

    Installing the G1330B Thermostat Transporting the Thermostatted Autosampler Transporting the Thermostatted Autosampler When moving the thermostatted autosampler around the laboratory, make sure that any condensed water inside the thermostat is removed. Tilt the module to the front, so that the water inside the thermostat can safely flow into the leak funnel.
  • Page 41: Troubleshooting And Test Functions

    Troubleshooting and Test Functions The thermostatted autosampler‘s built-in troubleshooting and test functions...
  • Page 42: Status Indicators

    Troubleshooting and Test Functions Status Indicators The thermostatted autosampler is provided with two status indicators which indicate the operational state (prerun, run, and error states) of the instrument. Both are located on the autosampler module. The status indicators provide a quick visual check of the operation of the thermostatted autosampler (see “Status Indicators”...
  • Page 43: Status Indicators

    Status Indicators Two status indicators are located on the front of the autosampler. The lower left indicates the power supply status, the upper right indicates the thermostatted autosampler status. Figure 18 Location of Status Indicators Status indicator Power supply indicator...
  • Page 44: Power Supply Indicator

    Troubleshooting and Test Functions Power Supply Indicator Power Supply Indicator The power supply indicator is integrated into the main power switch. When the indicator is illuminated (green) the power is ON.
  • Page 45: Instrument Status Indicator

    Instrument Status Indicator The instrument status indicator indicates one of four possible instrument conditions: • When the status indicator is OFF (and power switch light is on), the instrument is in a prerun condition, and is ready to begin an analysis. •...
  • Page 46: Als Thermostat Error Messages

    Troubleshooting and Test Functions ALS Thermostat Error Messages ALS Thermostat Error Messages Error messages are displayed in the user interface when an electronic failure occurs with the ALS thermostat module which requires attention. In the event of such a failure, the red status indicator at the front of the thermostatted autosampler is switched on, and an entry is written into the instrument log book.
  • Page 47: Fan Failed

    Troubleshooting and Test Functions Fan Failed Fan Failed Each fan in the ALS thermostat is equipped with a speed sensor, that allows monitoring and control the speed of the fans. If the sensor shows no signal, when the fan is activated, it is very likely that the fan is defective. The number in brackets indicates the fan position.
  • Page 48: Temperature Control Failed (1 - 4)

    Troubleshooting and Test Functions Temperature Control Failed (1 - 4) Temperature Control Failed (1 - 4) There are four Peltier elements build into the ALS thermostat for efficient cooling / heating. The electronics monitor the current through the Peltier elements. If the Peltier current is out of a specified limit the Peltier element is defective.
  • Page 49: Temperature Sensor Failed (1 - 4)

    Troubleshooting and Test Functions Temperature Sensor Failed (1 - 4) Temperature Sensor Failed (1 - 4) The ALS thermostat is equipped with four sensors to monitor the performance of the instrument. The sensors are connected to both sides of the Peltier elements for control of the cooling / heating efficiency. As the positioning of the sensor is critical to meet the performance requirements of the ALS thermostat the single sensors cannot be replaced separately.
  • Page 50: Lost Contact To Als Thermostat

    ❏ Check cable connection between autosampler and ALS thermostat Suggested Actions: ❏ Connect cable between autosampler and ALS thermostat ❏ Exchange ASM board in the autosampler ❏ Exchange TCA board in the thermostat ❏ contact the Agilent Technologies service organization...
  • Page 51: Power Fail For Als Thermostat Module

    • ALS thermostat module not connected to line power • ALS thermostat module fuse(s) defective • ALS Thermostat power supply defective ❏ Check correct power line connection Suggested Actions: ❏ Check fuse(s) and replace if defective ❏ contact the Agilent Technologies service organization...
  • Page 52: Temperature Out Of Range

    Troubleshooting and Test Functions Temperature out of Range Temperature out of Range To protect the Peltier elements for damage their working range is limited to - 3°C to 65°C. The error message is generated when at least one of the three Peltier elements exceeds this limit.
  • Page 53: Bad Cooling / Heating Performance

    10 min for the peltier elements to equilibrate ❏ Make sure that there is enough space on left and right side of the ALS thermostat and that the air channel inlets and oulets are not blocked ❏ contact the Agilent Technologies service organization...
  • Page 54 Troubleshooting and Test Functions Bad Cooling / Heating Performance...
  • Page 55: Repairing The Als Thermostat Module

    Repairing the ALS Thermostat Module Instructions on repair procedures of the ALS thermostat...
  • Page 56: Simple Repairs

    Repairing the ALS Thermostat Module Simple Repairs The ALS thermostat is designed for easy repair. Exchanging Internal Parts Some repairs may require exchange of defective internal parts. Exchange of these parts requires removing the ALS thermostat from the stack, removing the covers, and disassembling the ALS thermostat.
  • Page 57: Cleaning The Thermostat

    Repairing the ALS Thermostat Module Cleaning the Thermostat Cleaning the Thermostat The thermostat covers should be kept clean. Cleaning should be done with a soft cloth slightly dampened with water or a solution of water and a mild detergent. Do not use an excessively damp cloth that liquid can drip into the autosampler.
  • Page 58: Using The Esd Strap

    Repairing the ALS Thermostat Module Using the ESD Strap Using the ESD Strap Electronic boards are sensitive to electrostatic discharge (ESD). In order to prevent damage, always use an ESD strap supplied in the autosampler accessory kit, when handling electronic boards and components. Using the ESD Strap 1 Unwrap the first two folds of the band and wrap the exposed adhesive side firmly around your wrist.
  • Page 59: Overview

    Repairing the ALS Thermostat Module Overview Overview Figure 19 shows the main assemblies and their locations which can be repaired. Figure 19 Main assemblies Power Supply TCA Board Heat exchanger fan Heat exchanger Condensation leak drain...
  • Page 60: Exchanging The Power Supply Fuses

    Repairing the ALS Thermostat Module Exchanging the Power Supply Fuses Exchanging the Power Supply Fuses Frequency When defective Tools required Flat head screwdriver Parts required Fuses T2.5 A/250V (CSA, UL listed), 2110-0015 The fuse holders are located on the rear panel of the ALS thermostat. 1 Switch off the power switch at the front of the thermostatted autosampler.
  • Page 61: Removing The Top Cover And Foam

    Repairing the ALS Thermostat Module Removing the Top Cover and Foam Removing the Top Cover and Foam Tools required Screwdriver Pozidriv #1 Preparations for this Switch off ALS at the main power switch. Disconnect ALS and ALS thermostat power cords. procedure Remove ALS to ALS thermostat cable, and remove thermostat from stack.
  • Page 62 Repairing the ALS Thermostat Module Removing the Top Cover and Foam 3 Lift the clips on both sides of the top cover 4 Unscrew the screws on the top plate and (1). Remove the top cover (2). remove the plate by lifting its back first and then sliding to the front.
  • Page 63: Exchanging Heat Exchanger Fan

    Exchanging Heat Exchanger Fan When required When defective Removing the Top Cover and Foam (see “Removing the Top Cover and Foam” on page 61) Preparations for item 1 to 4. this procedure Fan, PN 3160-1079 Parts required 1 Unplug the connector at the TCA board and 2 Carefully lift the Fan upwards.
  • Page 64 Repairing the ALS Thermostat Module Exchanging Heat Exchanger Fan 5 Reinstall the fan and route the wire in the 6 Reconnect the wire to the connector. appropriate guide channel.
  • Page 65: Exchanging The Heatsink Fans

    Repairing the ALS Thermostat Module Exchanging the Heatsink Fans Exchanging the Heatsink Fans When required Fan not running Preparations for this Removing the Top Cover and Foam (see “Removing the Top Cover and Foam” on page 61). procedure Parts required Fan, part number 3160-0884 1 Carefully lift the pair of Fans upwards.
  • Page 66 Repairing the ALS Thermostat Module Exchanging the Heatsink Fans 3 Reinsert the Top Foam and route the cables in 4 Reconnect the connectors in the correct the guide channels in the Top Foam. order (See label on top of Power Supply). 5 Replace the metal cover and top cover.
  • Page 67: Exchanging The Sensors

    Repairing the ALS Thermostat Module Exchanging the Sensors Exchanging the Sensors When required Sensors defective Preparations for this Removing the Top Cover and Foam (see “Removing the Top Cover and Foam” on page 61). procedure Parts required Sensor flexboard, PN G1330-66504 Heatsink Compound, PN 6040-0454 1 Unscrew the six holding screws on top of the 2 Carefully lift the upper part of hardfoam...
  • Page 68 Repairing the ALS Thermostat Module Exchanging the Sensors 3 Remove the sealing adapter by carefully 4 Unplug the connectors at the TCA Board, sliding it out of the seat. careful lift the sensors out of their seat and remove the sensor flexboard. 5 Insert some Heatsink Compound into the four 6 Install the new sensor flexboard carefully by sensor seats, to assure good contact.
  • Page 69 Repairing the ALS Thermostat Module Exchanging the Sensors 7 Reinsert the sealing adapter by carefully 8 Reinstall the hardfoam encapsulation. sliding it in to the seat. 9 Route the cables in the guide channels and 10 Reinstall the Top Plate and the Top Cover. reconnect the connectors in the correct order (See label on Power Supply).
  • Page 70: Exchanging The Tca Board

    Repairing the ALS Thermostat Module Exchanging the TCA board Exchanging the TCA board When required TCA Board defective Preparations for this Removing the Top Cover and Foam (see “Removing the Top Cover and Foam” on page 61), procedure item 1 to 4. Parts required TCA board, part number G1330-66500 1 Remove all connectors on the TCA Board.
  • Page 71 Repairing the ALS Thermostat Module Exchanging the TCA board 3 By carefully pulling back on the upper part of 4 Remove the TCA board. the TCA board unsnatch it from the holder. 5 Reinsert the new TCA board. 6 Gently push forward on the upper part of the board, until it snatches into place.
  • Page 72 Repairing the ALS Thermostat Module Exchanging the TCA board 7 Refit the connector screws and reconnect all 8 Replace the top foam section, metal cover connectors. and top cover.
  • Page 73: Assembling The Main Cover

    Repairing the ALS Thermostat Module Assembling the Main Cover Assembling the Main Cover When required If cover is broken Tools required None Parts required Cover kit G1330-68703 (includes base, top, left and right) N O T E The cover kit contains all parts, but it is not assembled. WAR NI N G In case you insert the left or right side in the opposite position, you may not be able to remove the side from the top part.
  • Page 74 Repairing the ALS Thermostat Module Assembling the Main Cover...
  • Page 75: Als Thermostat Parts And Materials

    ALS Thermostat Parts and Materials Lists for identification of parts and materials...
  • Page 76: Main Assemblies

    ALS Thermostat Parts and Materials Main Assemblies Main Assemblies Figure 20 Main Assemblies Table 4 Main Assemblies Item Description Part Number Fuse - Power Supply (T2.5A/250V; CSA, UL listed) 2110-0015 Fuse TCA - Board (T3A/250V; CSA, UL listed) 2110-0029 Front Cover 5062-8582 Cable, autosampler - ALS thermostat G1330-81600...
  • Page 77: Accessory Kit G1330-68705

    ALS Thermostat Parts and Materials Accessory Kit G1330-68705 Accessory Kit G1330-68705 Table 5 Accessory Kit Item Description Part Number 5062-2463 Waste Tube Waste Tube Assembly G1330-67300 Reorder Number (5 m)
  • Page 78: Main Assemblies

    ALS Thermostat Parts and Materials Main Assemblies Main Assemblies Figure 21 Main Assemblies Table 6 Main Assemblies Item Description Part Number Heat sink fans 3160-0884 TCA Board G1330-66500 Heat exchanger fan 3160-1079...
  • Page 79: Foam Parts

    ALS Thermostat Parts and Materials Foam Parts Foam Parts Figure 22 Foam Parts Table 7 Main Assemblies Item Description Part Number Top Foam G1330-40102 Bottom Foam G1330-40103...
  • Page 80: Plastic Parts

    ALS Thermostat Parts and Materials Plastic Parts Plastic Parts Figure 23 Plastic Parts Table 8 Main Assemblies Item Description Part Number Cabinet Kit, includes base, top and sides G1330-68703 Front Cover 5062-8582 Leak Pan 5042-1307...
  • Page 81: Heat Exchanger Fan Parts

    ALS Thermostat Parts and Materials Heat Exchanger Fan Parts Heat Exchanger Fan Parts Figure 24 Heat Exchanger Fan Parts Table 9 Main Assemblies Item Description Part Number Fan Gasket G1330-44104 Heat Exchanger Fan 3160-1079...
  • Page 82 ALS Thermostat Parts and Materials Heat Exchanger Fan Parts...
  • Page 83: Specifications

    Specifications Performance specifications of the ALS thermostat...
  • Page 84: Performance Specifications

    Specifications Performance Specifications Performance Specifications Table 10 Performance Specifications Agilent 1100 Series thermostatted autosampler Type Specification Temperature range: setable from 4°C to 40°C in 1° increments...
  • Page 85 Index ESD strap 58 Agilent on internet 94 repairs 56 exchanging air circulation 21 damper 63 ALS thermostat accessory kit 24 fan 65, 67, 70 ALS thermostat accessory kit contents sample tray 10 internal parts 56 sample trays ALS thermostat operation 12 numbering of vial positions 36 screwdriver, pozidriv #1 65, 67, 70 ALS thermostat parts and materials...
  • Page 86 Index...
  • Page 87 Warranty Statement All Chemical Analysis Products Agilent Technologies warrants its chemical analysis products against defects in materials and workmanship. For details of the warranty period in your country, call Agilent. During the warranty period, Agilent will, at its option, repair or replace products which prove to be defective.
  • Page 88 Warranty Statement THE WARRANTY SET FORTH IS EXCLUSIVE AND NO OTHER WARRANTY, WHETHER WRITTEN OR ORAL, IS EXPRESSED OR IMPLIED. AGILENT SPECIFICALLY DISCLAIMS THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. Limitation of Remedies and Liability THE REMEDIES PROVIDED HEREIN ARE BUYER’S SOLE AND EXCLUSIVE REMEDIES. IN NO EVENT SHALL AGILENT BE LIABLE FOR DIRECT, INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES (INCLUDING LOSS OF PROFITS) WHETHER BASED ON CONTRACT, TORT OR ANY OTHER LEGAL THEORY.
  • Page 89 Warranty Statement Responsibilities of Agilent Technologies Agilent Technologies will provide warranty services as described in Table 1. Table 1 Warranty Services Services During Warranty Warranty Period Type Agilent 1100 Series of Modules 1 Year Onsite GC, LC, UV-Visible, and LAS supplies and...
  • Page 90: Safety Information

    Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards of design, manufacture, and intended use of the instrument. Agilent Technologies assumes no liability for the customer’s failure to comply with these requirements.
  • Page 91 Safety Information Capacitors inside the instrument may still be charged, even though the instrument has been disconnected from its source of supply. Dangerous voltages, capable of causing serious personal injury, are present in this instrument. Use extreme caution when handling, testing and adjusting.
  • Page 92 Safety Information WAR NI N G A warning alerts you to situations that could cause physical injury or damage to the equipment. Do not proceed beyond a warning until you have fully understood and met the indicated conditions. CA UT IO N A caution alerts you to situations that could cause a possible loss of data.
  • Page 93: Radio Interference

    Radio Interference Radio Interference Manufacturer’s Declaration This is to certify that this equipment is in accordance with the Radio Interference Requirements of Directive FTZ 1046/1984. The German Bundespost was notified that this equipment was put into circulation, the right to check the series for compliance with the requirements was granted.
  • Page 94: Agilent Technologies On Internet

    Agilent Technologies on Internet Agilent Technologies on Internet For the latest information on products and services visit our worldwide web site on the Internet at: http://www.agilent.com/go/chem...
  • Page 96 In This Book This manual contains information about the Agilent 1100 Series thermostatted for autosamplers. Information for the individual autosamplers can be obtained from the corresponding autosampler reference manual. This supplemental manual describes the following: • installing the thermostat, • theory of operation, •...

Table of Contents