Sequence
When you start the calibration, a wizard is started that guides you through calibration.
Depending on the type of calibration selected, the wizard includes the following steps:
Figure 7-4 Steps during calibration
7.7.5.2

Carrying out calibration/validation

Procedure
1. Call up submenu "5. Calibrations":
Component > Main menu > "1. Quick Start" > "5. Calibrations"
Depending on the wizard you enable, you either execute a calibration or validation. The
parameter assignment of this menu is used for both processes.
2. Select components to be calibrated/validated: "Components" parameter field.
– Authorize access with Standard PIN, if necessary.
– Enable/disable one or more component-related check boxes in the displayed editing
3. Specify calibration type: "Calibration type" parameter field.
– If you select "2-pt.", zero gas calibration and span gas calibration are executed one
Quick Start
Compact Operating Instructions, 05/2018, A5E31805656-06
window.
after the other. Otherwise you set either a zero gas calibration or a span gas
calibration.
Commissioning
7.7 Commissioning the device
99