Siemens SIPROCESS GA700 Quick Start Manual page 85

Continuous gas analysis
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

3. Wait for transition to warming-up phase:
When the splash screen disappears, the device is in the warming-up phase. A bar chart
in the dialog window shows the current temperature progress.
4. Access the measured value display during the warm-up phase:
– Close the warming-up window by pressing <ENTER>.
Setting up
Note
Operator control options
When commissioning for the first time, two methods can be used. You can commission the
device with the simplified process using the "[1] Quick Start" menu or opt for the more
detailed procedure described below.
• Additional information on the operating steps of the detailed procedure is available in the
operating instructions, see Table A-1 References 1 - Operating Manuals LUI (Page 135).
Make the necessary settings in the appropriate submenus.
• When you are working with the "[1] Quick Start" menu, you set the parameters of the
following menus one after the other:
→[1.1] Input of basic settings (Page 87)
→[1.2] Set automatic log-off (Page 88)
→[1.3] Set analog outputs (Page 89)
→[1.4] Measuring ranges (Page 91)
→[1.5] Calibration (Page 95)
→[1.6] Save parameter set (Page 101)
Short description of the detailed procedure
If you do not use the quick start procedure, commission the device with the steps described
below:
1. Call the main menu:
– Switch the measured-value display to selection mode. To do so, press <ENTER>.
– Press <Enter> again to open the main menu.
2. Identify device: → [3.09] Maintenance & Diagnostics > Identification.
3. Specify the PIN for the user levels:→ [5.1] Security > Access management.
In the factory settings, the PIN 1111 is the default for the "Standard" user level and PIN
2222 for the "Expert" user level.
Quick Start
Compact Operating Instructions, 05/2018, A5E31805656-06
The display shows the measured-value display in read mode.
For the duration of the warming-up phase, the device is in "Function check" mode.
The following symbol is displayed in the status bar:
Commissioning
7.7 Commissioning the device
85