OXYMAT 7
Note
Additional clamping screws
Additionally required clamping screws are not included in the scope of delivery. Order the
clamping screws separately.
Table 4- 1
Operating
mode
Pressure
operation
Suction ope-
ration
ULTRAMAT 7
Table 4- 2
Operating mode
Pressure operation/Suction operation
Quick Start
Compact Operating Instructions, 05/2018, A5E31805656-06
Mounting position of the sample gas restrictor: OXYMAT 7
Operation with one
OXYMAT 7
Install a sample gas re-
strictor in front of the sam-
ple gas inlet.
Install a sample gas re-
strictor in the sample gas
outlet.
Mounting position for the sample gas restrictor: ULTRAMAT 7
Sample gas restrictor/clamping screw
Operation with two
OXYMAT 7 in series
Install a sample gas re-
strictor in front of the sam-
ple gas inlet, into which the
sample gas is introduced
first.
In addition, replace the
sample gas restrictor in the
inlet of the rear, second
module with a clamping
screw.
Install a sample gas re-
strictor in the sample gas
outlet from which the sam-
ple gas is first sucked out.
In addition, replace the
sample gas restrictor in the
inlet of the rear, second
module with a clamping
screw.
Sample gas restrictor/clamping screw
Operation with one or two ULTRAMAT 7
No sample gas restrictor/clamping screw required
Combination operation
4.3 Gas connections
Operation with two
OXYMAT 7 in parallel
Install a sample gas re-
strictor in front of every
sample gas inlet.
Install a sample gas re-
strictor at every sample
gas outlet.
23