4. A separate Super Sani-Cloth should
be used for cleaning and
disinfecting. For disinfecting the
meter, get a new cloth and repeat
steps 2 and 3, making sure that the
surface of the meter stays wet for
2 minutes.
NOTE
Roche has tested the approved product for a total of 260 cycles which is equal to cleaning and
disinfecting once per week for a 5 year period.
5. Wash hands thoroughly with
soap and water.
87