Channel-To-Channel Isolated Analog Input Stage - National Instruments SCXI-1126 User Manual

Eight-channel isolated frequency input module for signal conditionin
Hide thumbs Also See for SCXI-1126:
Table of Contents

Advertisement

Channel-to-Channel Isolated Analog Input Stage

Note
If you are using Traditional NI-DAQ, the formula is V
© National Instruments Corporation
The SCXI-1126 consists of eight isolated frequency input channels, each
with software-programmable frequency ranges of 0–250 Hz, 0–500 Hz,
0–1 kHz, 0–2 kHz, 0–4 kHz, 0–8 kHz, 0–16 kHz, 0–32 kHz, 0–64 kHz,
and 0–128 kHz. Each channel has its own software-programmable input
threshold and hysteresis with a voltage range of –0.5 to 4.5 V. In addition,
each channel also has a 4-pole, lowpass output filter with
software-programmable cutoffs of 1 Hz, 40 Hz, 320 Hz, and 1 kHz.
The SCXI-1126 also has a digital section for automatic control of channel
scanning, frequency range selection, filter selection, and input threshold
and hysteresis level setting.
The input stage of each channel is isolated from the other channels and
from the chassis, up to 250 V
overvoltages, up to 250 V
This stage consists of a trigger circuit that compares the input waveform to
a user-programmed threshold and hysteresis. The programmable range of
the threshold is –0.5 to 4.48 V, and the programmable range of the
hysteresis is 0 to 4.98 V. You can extend these ranges by using an attenuator
terminal block. For example, the SCXI-1327 100:1 high-voltage attenuator
terminal block extends the threshold range from –50 to 250 V, and the
hysteresis range from 0 to 250 V. The threshold and hysteresis are
programmed to create a window with the upper and lower limits given by
the formula:
V
window
where (V
– V
threshold
hysteresis
When the input waveform crosses through this window, a trigger occurs.
The frequency of these triggers establishes the frequency to be converted
by the frequency-to-voltage conversion stage.
Chapter 1
Installing and Configuring the SCXI-1126
. Each channel is also protected from input
rms
powered on or off.
rms
(V
V
=
threshold
hysteresis
) must be greater than –0.5 V.
1-3
) to V
threshold
= V
±0.5 V
window
threshold
SCXI-1126 User Manual
.
hysteresis

Advertisement

Table of Contents
loading