Figure 11‑29 The message when the edited ID is not in the database
7.
Press [SAVE/SEND] with or without an active measurement in order for the
edited ID to become a permanent part of the database.
The sequence resumes at the previous active ID.
11.6.3

Erasing Data in the Active File

Data erased by the following techniques CANNOT be recovered.
To erase a single measurement
1.
From the measurement screen, press [ID#].
The ID review screen appears and shows the active ID with its stored data.
2.
Select the ID to be erased (see "Reviewing Stored Data and Changing the Active
ID" on page 199).
3.
Press [2nd F], [FILE] (CLR MEM) to erase the data at the displayed ID.
The displayed ID changes to the next ID in the sequence.
CAUTION
DMTA-10004-01EN, Rev. D, November 2016
Using the Datalogger
201