Binary Inputs (Function 72) - Siemens OXYMAT 64 Operating Instructions Manual

Gas analyzer for measurements of trace oxygen
Hide thumbs Also See for OXYMAT 64:
Table of Contents

Advertisement

Operation
5.4 Operating functions
Removing faults (Page 109)
5.4.6.4

Binary inputs (function 72)

Binary input functions
72 Binary inputs
Fault/Maint.r/CTRL NAMUR
Binary inputs define
Figure 5-49
Six floating binary inputs are included in the standard equipment ("0" = 0 V [0 to 4.5 V]; "1" =
24 V [13 to 33 V]).
If these six inputs are insufficient, you can retrofit additional electronics with eight further
binary inputs (optional board).
If you activate the "NAMUR" ("■") operating mode with softkey 1, the binary inputs behave as
in table "Binary input activation functions" marked with "N".
If you deactivate the "NAMUR" ("□") operating mode, the binary inputs behave compatibly
with the software outputs of the older version V 4.3.0 (in the table "Binary input activation
functions" marked with "X").
Press softkey 2. The following function screen appears on the display.
72 Binary inputs
B 1
B 2
B 3
B 4
Figure 5-50
Assign one of the activation functions listed below to each input as you like, whereby every
function may only be assigned once.
84
Binary input function
Vacant
Vacant
Vacant
Vacant
...cont.
Define binary inputs
O
2
O
2
Operating Instructions, 06/2008, A5E00880383-03
OXYMAT 64

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

7mb2041

Table of Contents