5.4.6
Configuring
5.4.6.1
Configuration submenu
Configuration
7O Analog outputs
71 Relay outputs
72 Binary inputs
73 ELAN - Configuration
Figure 5-46
All functions of this menu are only accessible via the code of level 2.
This menu contains all functions required for configuring the device.
Press the fifth softkey ("...continue") to branch to further configuration functions.
5.4.6.2
Analog output (function 70)
7O Analog output
Figure 5-47
With this function, you can configure the analog output.
With softkey 1, define the start-of-scale value of the analog current range. The following
settings (see also table "Configurations of the analog output") are possible:
● 0 – 20 mA
● 2 – 20 mA
● 4 – 20 mA
● NAMUR/4 – 20 mA (with limit at 3.8 mA).
OXYMAT 64
Operating Instructions, 06/2008, A5E00880383-03
...continue
Configuration submenu
4 - 2O mA
output inverted
Suppress negative
measured values
Configuring the analog output
O
2
O
2
Operation
5.4 Operating functions
81