Device Operating Modes - Siemens OXYMAT 64 Operating Instructions Manual

Gas analyzer for measurements of trace oxygen
Hide thumbs Also See for OXYMAT 64:
Table of Contents

Advertisement

5.1.4

Device operating modes

Table 5- 2
Mode
Coded display mode
Decoded display mode
Operator control mode
OXYMAT 64
Operating Instructions, 06/2008, A5E00880383-03
Operating modes
Properties
Measured value display is
shown
Protected submenus can
only be reached by entering
a code
The current operating state
of the device (except for
"Measure") is displayed in
the bottom line
Functional check not active
Measured value display is
shown
The submenus protected by
the entered code are
accessible
The current operating state
of the device (except for
"Measure") is displayed in
the bottom line
Functional check active
Measured value can be
influenced
Menu or function is
displayed
Settings and inputs can be
made
Functional check active
Measured value can be
influenced
5.1 General information about operation
Remarks
The device only supplies
reliable measured values in this
mode, as far as it's in the
"Measure" operating state. From
the operator control mode, you
can reach this mode by pressing
the [MEAS] button twice.
From the operator control mode,
you can reach this mode by
pressing the [MEAS] button
once and confirming or
discarding the made entries.
From "Coded display mode" you
can reach this mode by entering
the code of the corresponding
operation level.
In this mode, you can
parameterize/calibrate the
device.
Operation
45

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

7mb2041

Table of Contents