LOAD CELL TRIM
1: 2.941 mV
2: 3.334 mV
3: 3.148 mV
4: 1.747 mV
9. Place a test weight of at least 10% of the scale capacity on each corner of the scale, one
at a time, and note the reading for each corner.
10. Place a test weight on the corner with the lowest reading and adjust the appropriate cell
mV reading until the reading agrees with the highest reading obtained in Step 9.
Press the  key to increase the mV reading OR press the  key to decrease the mV
reading for each cell.
11. Repeat Step 10 until all corners have the same reading.
Load Cell Cable
12. Press the Exit key to exit Load Cell Trim and return to the Setup Page 5.
13. Press the Exit key to exit Setup and return to the Weight screen.
30
-
-
-
-
0.00
Exit
5.97
Cell locations are
for reference only!
+
+
+
+