4. The display will change to show "Working" for a few seconds and then change to show
"Set test weight over Cell 1".
CALIBRATE SCALE
Working
To calibrate zero,
remove any weight
from the scale
before proceeding
SAVE
5. Place the calibrated test weight on the scale over Cell 1 and then press the Save key.
CALIBRATE SCALE
Set test weight
over Cell 1 and
press Save.
SAVE
6. The display will change to show "Working" for a few seconds and then change to show
"Set test weight over Cell 2".
CALIBRATE SCALE
Set test weight
over Cell 2 and
press Save.
SAVE
7. Place the calibrated test weight on the scale over Cell 2 and then press the Save key.