8. With the "Known test weight" screen displayed, place the third calibrated test weight
(full test load) on the scale, enter the value for the test weight and then press the
Save key.
CALIBRATE SCALE
1000.0
This is the known
test weight
that is currently
on the scale.
SAVE
9. The display will change to show "Working" for a few seconds and then return to
Setup Menu Page 5.
10. Calibration is now complete.
 Press the Exit key to exit Setup and return to the Weight screen.
28
1
4
7
0
CANCEL
2
3
5
6
8
9
.
DEL
NOTE: A minimum of 50%
of the scale's capacity is
required for the full test
load, however 70% to 100%
is recommended.
3300-0298-0M Rev B 05/18