Detecto Solace ID-3636S-855RMP Manual page 11

In-floor dialysis scale
Hide thumbs Also See for Solace ID-3636S-855RMP:
Table of Contents

Advertisement

Setup Menu Page 1
MENU
Cal=9, Cfg=5
Units: lb, in
Auto Locking: 4 sec.
Sonar Ht: 222.2 cm
<<<
Cal=XX, Cfg=YY – Audit Trail Counters
These are the Audit Trail Counters for the number of times that the metrological and non-
metrological settings have been changed. This is for information purposes only.
Units:
Touching this key will toggle the weighing and height units between (lb, in) or (kg, cm).
Note that the units cannot be toggled if OIML is enabled.
NOTE: When you calibrate the indicator, ensure that the proper base units are selected
here before attempting to calibrate. It is safe to change the units without calibrating, just
be sure not to change any other metrological settings when you are using the converted
units as this may place your calibration to an invalid state.
Allowable settings are lb, in (pounds, inches) or kg, cm (kilograms, centimeters)
Auto Locking:
This setting is used by the indicator to hold a stable patient weight for a desired amount of
time. For example, if a value of 5 seconds is used, then when the indicator locks onto a
stable patient weight, it will remain locked for 5 seconds before automatically releasing the
weight.
Touching this key will open the entry screen for the auto-locking feature of the indicator. The
value entered is a time in seconds. Using the numeric keys, enter a new value and then
press the Save key to save it.
Allowable values for auto locking are 0 to 60.
Sonar Ht:
This is the height in centimeters from the scale platform to the bottom of the sonar sensor
enclosure. This value is used to determine the patient's height and should be measured and
entered as accurately as possible to guarantee accurate height measurements using the
sonar height rod.
Touching this key will open the entry screen for the sonar sensor height. The value entered
is in centimeters. Using the numeric keys, enter a new value and then press the Save key to
save it.
 Press the >>> key to advance to Setup Menu Page 2.
 Press the <<< key to return to the Setup Review screen.
 Press the Exit key to exit Setup and return to the Weight screen.
10
PAGE 1
Exit
>>>
3300-0298-0M Rev B 05/18

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Solace id-4848s-855rmpSolace id-3636sh-855rmpSolace id-4848sh-855rmpSolace id-4836s-855rmpSolace id-7248s-855rmpSolace id-4836sh-855rmp ... Show all

Table of Contents