National Instruments SCXI-1305 Installation Manual page 11

Ac/dc coupling bnc terminal block
Hide thumbs Also See for SCXI-1305:
Table of Contents

Advertisement

Mechanical
Environmental
Safety
© National Instruments Corporation
Maximum working voltage
(signal + common-mode)
Channel to ground........................... Each channel must remain within
Channel to channel.......................... Each channel must remain
Field-wiring connectors
Signal connectors ............................ 8 female BNC connectors
Auxiliary connectors....................... 1 female SMB connector
Signal connector spacing ................ 1.8 cm (0.75 in.)
Ground-referencing
You can configure the SCXI-1305 for use with either
ground-referenced or non-referenced (floating) signal sources on a
per-channel basis using slide switches. When a channel is configured
for use with a non-referenced (floating) signal source, CH– is
connected to analog ground through a 100 kΩ resistor.
Dimensions............................................. 17.8 by 9.7 by 3.0 cm
Weight .................................................... 200 gm
Operating temperature............................ 0 to 50 °C
Storage temperature ............................... –20 to 70 °C
Relative humidity ................................... 10 to 90% noncondensing
Altitude (maximum)............................... 2000 m
Designed in accordance with IEC 61010-1, UL 3111-1, and CAN/CSA
C22.2 No. 1010.1 for electrical measuring and test equipment
Installation Category II
Pollution degree 2
42 V
within 42 V
of the voltage applied to any
other channel
(7.0 by 3.125 by 1.2 in.)
(7 oz.)
11
or ±42 VDC of ground
rms
or ±42 VDC
rms
SCXI-1305 Installation Guide