    Summary Summary Technical Specifications Manufacturer Declaration of Conformity Purpose Specifications Labelling Labelling details Symbol description Contents of the package How to use Warning Electromagnetic Interference Contraindications How to use TENS and ionophoresis Programs list Maintenance Battery charging Battery replacement Cleaning Carriage and storage Disposal Troubleshooting...
  • Page 4: Technical Specifications

    Equipment not suited to be used in external. Purpose Clinical purpose : Therapeutic Use: Clinic/Hospital and domestic use MIO-IONOTENS is indicated for the treatment and the functional rehabilitation of the following pathologies and anatomical zones: IACER S.r.l. 4 di 28 MNPG151 Rev. 05 11/11/15...
  • Page 5: Specifications

     tendinitis MIO-IONOTENS, thanks to its protocols TENS, is particularly suitable for the treatment of pain. TENS pulses are able to significantly reduce, and in some cases eliminate, the sensation of pain caused by diseases and / or problems indicated above.
  • Page 6: Labelling

    Labelling CH1 output CH2 output Battery charger connector Display Increase intensity CH1 Decrease intensity CH1 Mode operation button Increase intensity CH2 Decrease intensity CH2 (10) Increase program (11) Decrease program (12) ON/OFF and OK button (13) Set programs and therapy pause button (14) Belt clip (15) Battery compartment IACER S.r.l.
  • Page 7: Labelling Details

    Mode operation (REHA, TENS, MEM) Wave frequency Wave impulse width Program number CH1 intensity CH2 intensity Battery status Therapy time Contraction time (10) Recovery time (11) Up/down slope Labelling details IACER S.r.l. 7 di 28 MNPG151 Rev. 05 11/11/15...
  • Page 8: Symbol Description

    Check the device and its accessories before use. Avoid the use in case of damage to the case or to the accessories (damaged cables): contact the manufacturer as mentioned in “Assistance” paragraph;  Avoid the use of MIO-IONOTENS to people not educated through the reading of the manual;  Avoid the use of MIO-IONOTENS in damp environments;...
  • Page 9: Electromagnetic Interference

    MIO-CARE PRO recognize the electrodes connection: in case of faulty connection, when the intensity reaches 10 mA the value is resetted to zero. The remaining time is showed on the display of MIO-IONOTENS. An acoustic signal advises the user when the treatment is completed.
  • Page 10: Tens And Ionophoresis

    The device automatically switches off when no button is pressed for 2 minutes. FREE MEMORIES (ADJUSTABLE PROGRAMS) With I-TECH PHYSIO you can set the parameters according to your needs using the MEM programs. Read the following instructions to adjust the parameters.
  • Page 11 Position the electrode with the drug on painful area, and the other electrode on the other side. At the end of the program, the skin could lightly turn bright red; the reddening usually vanishing few minutes after the end of program. WARNING.
  • Page 12: Programs List

    Lysine salt Thiomucase Negative Antiedemic Post-trauma and post- surgical oedema due venous insufficiency If the drug used is not included in the above list, determine the polarity from the package or consult the prescribing doctor or dispensing pharmacist. Programs list TENS REHA Conventional Tens...
  • Page 13 Programs Technical Specifications TENS Programs Prg. PHASE 1 PHASE 2 PHASE 3 Total time 40 min frequency 90 Hz impulse width 50s Time tot 30 min frequency 1 Hz impulse width 200s Time tot 3 min frequency 150 Hz impulse width 200s Total time 30 min frequency 120 Hz impulse width 40s...
  • Page 14 TENS 1 • Conventional TENS Program used for analgesic purposes; its purpose is to induce the organism into blocking pain at the spine, in accordance with the “Gate Control Theory” by Melzack and Wall. Pain impulses leave part of the body (for example the hand) and run along the nerve tracts (through small-diameter nerve fibers) until they reach the central nervous system where the impulses are interpreted as pain.
  • Page 15 TENS 5 • Neck pain / Headache Specific program for the treatment of pain in the neck area. The intensity should be adjusted to a level between the thresholds of perception and pain: the maximum intensity level is the moment in which the muscles surrounding the area treated begin to contract;...
  • Page 16 types of square-wave impulses has a general analgesic effect on the area to be treated (impulses at different frequencies stimulate different sized nerve fibres promoting an inhibitory action at spinal level). The intensity should be adjusted to a level between the thresholds of perception and pain, without causing muscle contraction: Position of electrodes: form a square above the area to be treated as shown in illustration 1.
  • Page 17 increasingly over time. Arthrosis mainly causes progressive deterioration of cartilage (which is not capable of re-forming) and bone, with secondary deformation of the same, and production of excrescences, called "osteophytes", which mechanically obstruct joint movement; it also causes the joint capsule to thicken and stiffen, which together with contraction of the muscles surrounding the joint limits the "joint excursion"...
  • Page 18 exception of the “periarthritis” program, the muscles surrounding the area to be treated must not contract, they should only produce slight “vibrations”. REHA Programs Prg. PHASE 1 PHASE 2 PHASE 3 Total time 30 min frequency 800 Hz impulse width 100s Total time 30 min frequency 1000 Hz impulse width 100s...
  • Page 19 REHA 4 • Microcurrent The use of microcurrent is very similar to conventional Tens, the only difference being the very fine electric impulse used that is sometimes more suitable for the sensibility of slightly anxious people or the more delicate parts of the body. It can generally be applied for everyday pains, bearing in mind that you should always consult your doctor to identify the cause of the pain.
  • Page 20: Maintenance

    Program to be carried out when loss of muscle tone has already occurred. Apply with caution (at low intensity, enough to produce light muscle contractions) in the first 2/3 weeks. Increase intensity progressively over the next 3/4 weeks. Application on alternate days. Position of electrodes from photo 1 to photo 20.
  • Page 21: Battery Replacement

    Carriage and storage Precautions for transportation MIO-IONOTENS is a portable device, so it does not need any particular carriage precautions. However we recommend to put away MIO-IONOTENS and its accessories in their own bag after every treatment. Storage precautions MIO-IONOTENS is protected till following environmental conditions:...
  • Page 22: Troubleshooting

    Troubleshooting If it is used in accordance with the instructions of the user manual, MIO-IONOTENS does not need a particular regular maintenance. If you find any malfunctioning using MIO-IONOTENS, please follow these instructions:  MIO-IONOTENS does not turn on and/or the display does not light up. Check the battery status and replace it if it is necessary (make reference to chapter "Battery replacement").
  • Page 23: Emc Tabels

    EMC Tabels Electromagnetic emission Electromagnetic environment – Emission test Compliance guidance The device uses RF energy only for its internal Group 1 function. Therefore, its RF emissions are very low RF emissions and are not likely to cause any interference in Cispr 11 nearby electronic equipment Class B...
  • Page 24 Rated maximum Separation distance according to the frequency of the transmitter (m) power transmitter (W) 150kHz to 80MHz 80MHz to 800MHz 800MHz to 2GHz d = 1,2 P d = 1,2 P d = 2,3 P 0,01 0,12 0,12 0,23 0,38 0,38 0,73...