Multichassis System - National Instruments SCXI Series Quick Start Manual

Signal conditioning extensions for instrumentation
Hide thumbs Also See for SCXI Series:
Table of Contents

Advertisement

Available languages

Available languages

3.
Fasten the adapter to the rear of the SCXI chassis with the screws provided with the SCXI-1349.
1
SCXI Chassis
2
SCXI-1349 Cable Adapter

Multichassis System

The SCXI-1346 covers the rear connector of two modules. When viewing the chassis from the rear, the
module to the right of the module directly connected to the SCXI-1346 cannot have an external cable
inserted into its rear 50-pin connector.
SCXI-1000 chassis through revision D do not have address jumpers or switches and respond to any
address, but you cannot use them in multichassis systems. Revision E chassis use jumpers on Slot 0 for
chassis addressing. Revision F and later chassis use a DIP switch for chassis addressing.
SCXI-1000DC chassis through revision C do not have address jumpers or switches and respond to any
address, but you cannot use them in multichassis systems. Revision D and later chassis use jumpers on
Slot 0 for chassis addressing.
SCXI-1001 chassis through revision D use jumpers on Slot 0 for chassis addressing. Revision E and later
chassis use a DIP switch for chassis addressing.
To connect the multichassis system, you must use one SCXI-1346 multichassis adapter for every chassis
in the chain except the chassis furthest from the DAQ communicating device. The last chassis uses the
SCXI-1349 cable adapter.
1.
Identify the appropriate SCXI module to connect to the cable adapter. Refer to step 1 of the
previous
2.
Insert the 50-pin female connection on the rear of the cable adapter into the 50-pin male connector
on the rear of the appropriate SCXI module.
3.
Fasten the adapter to the rear of the SCXI chassis with the screws provided with the SCXI-1346.
4.
Repeat steps 1 through 3 for each SCXI chassis in the system, excluding the last SCXI chassis in
the chain.
SCXI Quick Start Guide
Figure 10. Installing the Cable Adapter
Single-Chassis System
section to determine the appropriate module.
1
S E
FU
LL
2
P U
4
3
4
3
68-Pin Shielded Cable
4
Screws
8
ni.com