Troubleshooting and Maintenance

Troubleshooting

This section lists probable corrections for common hardware problems. Please check all of the
following items before requesting repairs.
Connection Errors
Problem
The POWER indicator is
not lit.
The Monitor is blank.
The Monitor image is not
clear.
Cannot make key inputs
from the Console.
Camera images do not
appear on the screen
(for Cameras with Light
Source)
Camera images do not
appear on the screen
(when a standard CCTV
lens and lighting are used)
F250
50
Setup Manual
Probable cause
The Power Supply is not connected properly.
The supply voltage is not 24 VDC +10%/-15%.
The power to the Monitor is not ON.
The Monitor Cable is not connected properly.
The Monitor is malfunctioning.
When using an Monitor, the power supply capacity is
insufficient.
There is electrical noise entering from the power supply or
cables.
The Monitor Cable is not connected properly.
The Console Cable is not correctly connected.
The Camera Cable is not correctly connected.
The lighting cable is not properly connected to the
Camera.
The lens cap has not been removed.
The Camera Cable is not correctly connected.
The lens iris is opened or closed too far.
The shutter speed is not suitable.
The lighting method is not suitable.
Reference
p.28
p.25
p.25
p.25
p.25
p.32