13 Maintaining the NPD
To Install a Capillary Column in the NPD
5 Break off the column end by supporting it against the column cutter opposite the scribe.
6 Wipe the column walls with a tissue dampened with isopropanol to remove fingerprints
7 Install the capillary column.
214
Inspect the end with a magnifying loupe to make certain there are no burrs or jagged
edges.
and dust.
If the column id is greater than 0.1 mm:
a Gently insert the column into the detector until it bottoms; do not attempt to force it
further.
b Finger-tighten the column nut, then withdraw the column about 2 mm. Tighten the nut
an additional 1/4 turn with a wrench.
Bad
Good
Maintaining Your GC