Agilent Technologies TwisTorr 305 FS Series User Manual page 241

Hide thumbs Also See for TwisTorr 305 FS Series:
Table of Contents

Advertisement

Two 6 mm (1/4") internal diameter rubber or plastic hoses from the water supply
must be fitted to the two nozzles.
These hoses must be held on the respective nozzles using clamps to avoid the
NOTE
tubes from becoming loose or disconnected during operation.
Cooling may be carried out either through an open circuit with eventual discharge of
the water, or using a closed circuit cooling system.
The water temperature must be between +15 °C and +30 °C with a flow rate higher
than 50 l/h (0.22 GPM). The max water inlet pressure must be lower than 5 bar.
The water electrical conductance must be ≤ 500 µs/cm. When the conductivity
NOTE
is higher, in closed water circuit, the use of up to 20 % of Ethyl-Glycole is
recommended.
TwisTorr 305 FS User Manual / 87-901-057-01
Water Cooling Kit Connection (only for water
Technical Information
cooling pumps)
241/258