[SENSe:]VOLTage[:DC]:RANGe[:UPPer]

Syntax
[SENSe:]VOLTage[:DC]:RANGe[:UPPer] {<range>|MIN|MAX|DEF}
This command selects the measurement range for DC voltage measurements.
[SENSe:]VOLTage[:DC]:RANGe[:UPPer]? [{MIN|MAX}]
This query returns a numeric value that represents the DC voltage
measurement range.
Parameter
range
Remarks
• Selecting a discrete range will disable autoranging.
• If the input signal is greater than can be measured on the selected range
• If a range change is in progress (due to the autoranging setting) when you
• The instrument is set to 100 mV, with autoranging enabled, after a Factory
U3606B Programmer's Reference
Item
Type
Numeric
(manual ranging), the instrument gives an overload indication.
Positive overload: "+9.900000E+37" or "OL" from the front panel.
Negative overload: "-9.900000E+37" or "–OL" from the front panel.
query the instrument, the value "+9.910000E+37" will be returned. Wait
for the instrument to select an appropriate range before querying the
instrument again.
Reset (*RST command).

[SENSe:]VOLTage[:DC]:RANGe[:UPPer]

Range of values
• 20 mV|MIN
• 100 mV
• 1 V
• 10 V
• 100 V
• 1000 V|MAX
• DEF (AUTO)
SENSe Subsystem
10
Default value
MIN