4
CONFigure Subsystem

CONFigure:LRESistance

CONFigure:LRESistance

Syntax
CONFigure:LRESistance
This command first resets all low-resistance measurement parameters and
trigger parameters to their default values. Then, it configures the instrument
for 4-wire low-resistance measurements but does not initiate the
measurement.
N O T E
Parameters
range
resolution
104
[<range>|AUTO|MAX|MIN|DEF[,{<resolution>|MAX|MIN|DEF}]]
The
CONFigure:LRESistance
in the "wait-for-trigger" state. Use the
READ?
place the instrument in the "wait-for-trigger" state.
Item
Type
Numeric
Numeric
command in conjunction with the
Range of values
• 100 mΩ|MIN
• 1000 mΩ
• 10 Ω
• 100 Ω
• 1000 Ω|MAX
• AUTO|DEF
• MAX (4½ digit)
10 uΩ
100 uΩ
1 mΩ
10 mΩ
100 mΩ
command does not place the instrument
INITiate[:IMMediate]
CONFigure:LRESistance
Default value
AUTO
• MIN (5½ digit)
MIN
1 uΩ
10 uΩ
100 uΩ
1 mΩ
10 mΩ
U3606B Programmer's Reference