Invacare® Soft Tilt
End user turning from right to left side, never laying in
a flat position and always laying on a flat wing.
– Angle primary wing = 13°-30°
– Angle secondary wing = 0°
– Time flat position = 0 min.
– Time left and right position > 0 min.
– see section Auto program sequence as an example.
End user turning from right to left side, never laying in
a flat position and always laying on a slightly elevated
wing (Cradle function).
– Angle primary wing = 13°-30°
– Angle secondary wing = 1°-12°
– Time flat position = 0 min.
– Time left and right position > 0 min.
This Program requires "high" side rails.
End user turning from one side to flat and back to the
same side, always laying on a flat wing.
– Angle primary wing = 13°-30°
– Angle secondary wing = 0°
– Time flat position > 0 min.
– Time left or right side position = 0 min.
16
End user turning from one side to flat and back to the
same side, always laying on a slightly elevated wing.
– Angle primary wing = 13°-30°
– Angle secondary wing = 1°-12°
– Time flat position > 0 min.
– Time left or right side position = 0 min.
This Program requires "high" side rails.
Both wings raised to equal angles without repositioning
(Backward hug).
– Angle primary and secondary wing = 10°-12°
– Time flat position = 0 min.
– Time left and right position > 0 min (set values not
relevant, see below)
No automatic repositioning will occur when both
wings are set to same angle and therefore, the
times set for left or right side position do not
have any influence.