SATO CL4NX Plus Operator's Manual page 225

Hide thumbs Also See for CL4NX Plus:
Table of Contents

Advertisement

[Auto-calibration]
Perform the auto-calibration for the selected media sensor.
[Auto-calibration] is not available if you have installed the optional linerless cutter kit.
Gap + I-Mark
Adjusts both the Gap sensor and I-mark sensor.
Gap
Adjusts the Gap sensor.
I-Mark
Adjusts the I-mark sensor.
The procedure to auto-calibrate the media sensor is as follows:
1. Push the head lock lever towards the rear to unlock the print head.
2. Pass the media below the media sensor guide. If you are using labels, remove the label from the
liner. Align it so that the media sensor does not detect the I-mark (black mark).
3. Press the print head down until the head lock lever is locked. To get the correct adjustment result,
adjust after you have locked the print head.
4. Press the
5. Press the
6. When the confirmation screen appears, press the
7. The result of the sensor adjustment is shown. To exit the adjustment, press the
8. Set to Offline mode. Press the
If the media is not fed correctly after doing [Auto-calibration], contact your SATO
reseller or technical support center.
/