UNI-T UTG1000A User Manual page 54

Dds function generator
Hide thumbs Also See for UTG1000A:
Table of Contents

Advertisement

After enabling FSK function, the default of hop frequency is 2MHz. If need to change, press
Parameter→Hop Frequency in turn. Settable range of hop frequency is determined by carrier wave
waveform. See the following table for setting range of each carrier wave frequency:
Frequency
Carrier
UTG1010A
Wave
Minimum
Waveform
Value
Sine Wave 1μHz
Square
1μHz
Wave
Ramp
1μHz
Wave
Arbitrary
1μHz
Wave
FSK Rate Setting
When modulation source is internal, the moving frequency between carrier wave frequency and hop
frequency can be set. After enabling FSK function, FSK rate can be set and the settable range is 2mHz to
100kHz, the default rate is 1kHz. If need to change, press Carrier Wave Parameter→Rate in turn.
Comprehensive Example
Firstly, make the instrument work in frequency shift keying (FSK) mode, then set a sine wave with 2kHz
and 1Vpp from the internal of the instrument as a carrier wave signal, and set hop frequency to 800 Hz,
finally, make carrier wave frequency and hop frequency move between each other with 200Hz
frequency. Specific steps are seen as following:
1) Enable Frequency Shift Keying (FSK) Function
Press Menu→Modulation→Type→Frequency Shift Keying in turn to start the FSK function.
UTG1005A
Maximum
Minimum
Value
Value
10MHz
1μHz
5MHz
1μHz
400kHz
1μHz
2MHz
1μHz
53
UTG1000A Series Function Generator User Manual
Maximum
Value
5MHz
5MHz
400KHz
1MHz