UTG1000A Series Function Generator User Manual
When modulation source is internal, frequency of ASK amplitude jump can be modulated. After
enabling ASK function, ASK rate can be set and the settable range is 2mHz to 100kHz, the default rate is
1kHz. If need to change, press Carrier Wave Parameter→Rate in turn.
Comprehensive Example
Make the instrument work in amplitude shift keying (ASK) mode, then set a logic signal with 300Hz
from the internal of the instrument as a modulation signal and a sine wave with frequency of 15kHz
and amplitude of 2Vpp as a carrier wave signal. Specific steps are seen as following:
1) Enable Amplitude Shift Keying (ASK) Function
Press Menu→Modulation→Type→Amplitude Shift Keying in turn to start the ASK function.
2) Set Carrier Wave Signal Parameter
Press Carrier Wave Parameter→Type→Sine Wave in turn
Press Parameter softkey, and the interface will pop up as following: