Definition of Terms - Visual Fault Finder Submodule

Output Power Level (CW)
Center Wavelength
206
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
The output power at the specified wavelength, measured at the end
of a jumper cable.
The wavelength representing the center of mass of selected peaks.
The power and wavelength of each used to calculate the mean
wavelength λ:
λ = Σ P
λ
/ Σ P
i
i
1
where: P
is the power of a single peak.
i
Characteristics
Horizontal Parameters
• Start-km: 0 km to 400 km
• Span: 0.1 km to 400 km
• Readout resolution: 0.1 m
• Minimum sample spacing: 8 cm
• Refractive index: 1.00000 to 2.00000
• Length unit: km, ft, or miles
• Measurement points: up to 16000
Vertical Parameters
• Vertical scale: 0.1 to 10.0 dB/Div
• Read-out resolution: 0.001 dB
• Reflectance range: -14 dB to -60dB
• Backscatter coefficient: 10 to 70 dB at 1
E6000C Mini-OTDR User's Guide, E0302
Characteristics
µs