Overload Protection - Siemens SIWAREX WL270 CP-S SB Operating Instructions Manual

Pressure piece set
Hide thumbs Also See for SIWAREX WL270 CP-S SB:
Table of Contents

Advertisement

Height compensation of the support points
With this method, the output signals of all of the load cells are attuned to each other under a
load. To ensure that all of the load cells receive approximately the same load, spacer plates
are used to compensate the heights.
The output signals of the individual load cells correspond to the load distribution.
Proceed as follows to measure the output signals:
1. Disconnect the load cell wires SIG+ and SIG-.
2. Supply the load cells with power from the power supply.
3. Measure the output voltage between SIG+ and SIG- of each individual load cell.
4. Place spacing plates under the load cell bracket with the lowest value until the output
4.4

Overload protection

CAUTION
Irreparable faults and damage to the load cells
When using load cells with small rated loads, always provide overload protection in order to
protect the cells from becoming damaged. This applies both to the measuring direction and
the transverse direction.
If load cells are used beyond the maximum working load or the maximum transverse load,
this can cause irreparable faults and even fracturing of the load cell or the installation unit.
Overload protection in the measuring direction
Load cells can be protected against vertical overload by limiting the measuring path with
stops.
For load cells with high rated loads, the risk of an unintentional overload is not as great.
Possible additional loads can already be taken into consideration during the dimensioning of
the load cells.
Another method for protecting load cells from an overload is to over-dimension the load cells.
CP-S SB
Operating Instructions, 11/2010, A5E03323543B
voltages are equal.
Application planning
4.4 Overload protection