Section 3: Mounting, Wiring Switches, and Sensors:
Models 702DX22 and 702DX61
February 2013
3.3

Wiring Switches and Sensors

Wiring Switches and Sensors with Dry Contact Inputs,
Measurement Option Code 22 (702DX22)
Figure 3-5. Rosemount 702 Terminal Diagram
3.3.1

Wireless Output Specifications

Dual Input
The 702 transmitter will accept the input from one or two single pole single throw switches on
inputs S1 and S2. The wireless output of the transmitter will be both a primary variable (PV) and
a secondary variable (SV). The PV is determined by the S1 input. The SV is determined by the S2
input. A closed switch drives a TRUE output. An Open switch drives a FALSE output.
Figure 3-6. Single, Dual Input
Single Input
S1
CMN
22
1
2
3
4
S2
CMN
Dual Input
S1
CMN
S2
CMN