Exchanging The Electronics Module - Siemens SITRANS LG260 Operating Instructions Manual

Guided wave radar. four-wire modbus converter version in second chamber rod and cable probe
Hide thumbs Also See for SITRANS LG260:
Table of Contents

Advertisement

Measurement error during emptying
Fault description
7. Measured value re-
mains unchanged in
the close range during
emptying
8. Measured value re-
mains reproducible
in one position during
emptying
Reaction after fault recti-
fication
56
Error pattern
Cause
False signal larger than the level
echo
Level echo too small
0
time
Stored false signals in this
position are larger than the level
echo
0
time
Depending on the reason for the fault and the measures taken, the
steps described in chapter "Setup" must be carried out again or must
be checked for plausibility and completeness.
8.5

Exchanging the electronics module

If the electronics module is defective, it can be replaced by the user.
In Ex applications, only instruments and electronics modules with ap-
propriate Ex approval may be used.
If there is no electronics module available on site, the electronics
module can be ordered through the agency serving you. The electron-
ics modules are adapted to the respective sensor and differ in signal
output or voltage supply.
The new electronics module must be loaded with the default settings
of the sensor. These are the options:
In the factory
Or on site by the user
In both cases, the serial number of the sensor is needed. The serial
numbers are stated on the type label of the instrument, on the inside
of the housing as well as on the delivery note.
When loading on site, the order data must first be downloaded from
the Internet (see operating instructions manual "Electronics module").
Caution:
All application-specific settings must be entered again. That's why you
have to carry out a fresh setup after exchanging the electronics.
If you saved the parameter settings during the first setup of the sen-
sor, you can transfer them to the replacement electronics module. A
fresh setup is then not necessary.
SITRANS LG260 - Operating Instructions
Rectification
Eliminate false signals in the
close range
Remove contamination on the
probe. After having removed the
source of the false signals, the
false signal suppression must
be deleted.
Carry out a new false signal
suppression
Delete false signal suppression
Carry out a new false signal
suppression
PBD-51041059