Siemens SITRANS LG260 Operating Instructions Manual page 74

Guided wave radar. four-wire modbus converter version in second chamber rod and cable probe
Hide thumbs Also See for SITRANS LG260:
Table of Contents

Advertisement

Register Name
Register Number
Address
200
Baud Rate
201
Parity
202
Stopbits
203
Delay Time
206
Byte Oder (Float-
3000
ing point format)
Input register
The input registers consist of 16 bits. They can only be read out. Before each command, the ad-
dress (1 byte) is sent, after each command a CRC (2 bytes) is sent.
PV, SV, TV and QV can be adjusted via the sensor DTM.
Register Name
Register Number
Status
100
PV Unit
104
PV
106
SV Unit
108
SV
110
TV Unit
112
TV
114
QV Unit
116
QV
118
Status
1300
PV
1302
SV
1304
TV
1306
QV
1308
Status
1400
PV
1402
Status
1412
SV
1414
74
Type
Configurable Values
Word
1 ... 255
Word
1200, 2400, 4800, 9600,
19200
Word
0 = None, 1 = Odd, 2 =
Even
Word
1 = None, 2 = Two
Word
10 ... 250
Word
0, 1, 2, 3
Type
Note
DWord
Bit 0: Invalid Measurement Value PV
Bit 1: Invalid Measurement Value SV
Bit 2: Invalid Measurement Value TV
Bit 3: Invalid Measurement Value QV
DWord
Unit Code
Primary Variable in Byte Order CDAB
DWord
Unit Code
Secondary Variable in Byte Order CDAB
DWord
Unit Code
Third Variable in Byte Order CDAB
DWord
Unit Code
Quarternary Variable in Byte Order CDAB
DWord
See Register 100
Primary Variable in Byte Order of Register 3000
Secondary Variable in Byte Order of Register 3000
Third Variable in Byte Order of Register 3000
Quarternary Variable in Byte Order of Register 3000
DWord
See Register 100
Primary Variable in Byte Order CDAB
DWord
See Register 100
Secondary Variable in Byte Order CDAB
SITRANS LG260 - Operating Instructions
Default
Unit
Value
246
9600
0
1
50
ms
0
PBD-51041059