9 Calibration
Reference Resistor Calibration
you press Exit the procedure will resume measuring resistance as in Step
5 above.
8.
Verify the calibration by having the 1575/1590 measure the resistance at-
tached to Input 2 using the 100Ω internal reference resistor. It should be
accurate within the specifications.
Repeat the procedure for each of the other reference resistors: 1Ω, 10Ω,
and 10 kΩ. When calibrating the 1Ω, 10Ω, or 10 kΩ reference resistors,
replace 100Ω in steps 1, 2, 4, and 8 with the respective reference resistor
value.