Table 12 Interface Commands Continued - Fluke Hart Scientific 1575 Super-Thermometer User Manual

Thermometer readout
Table of Contents

Advertisement

1575/1590 Thermometer Readout
User's Guide

Table 12 Interface Commands continued

Command
po[ffset](p)[=n]
rl[90](p)[=n]
rh[90](p)[=n]
r9[0](p)[=n]
a9[0](p,r)[=n]
b9[0](p,r)[=n]
c9[0](p,r[,i])[=n]
d9[0](p)[=n]
r6[8](p)[=n]
al6[8](p)[=n]
de6[8](p)[=n]
a6[8](p,r)[=n]
b6[8](p,r)[=n]
c6[8](p,r)[=n]
d6[8](p,r)[=n]
scal[e](p)[=9[0]/6[8]]
rc[vd](p)[=n]
alc[vd](p)[=n]
dec[vd](p)[=n]
bec[vd](p)[=n]
ap[oly](p,i)[=n]
at[herm](p,i)[=n]
bt[herm](p,i)[=n]
Display Commands
drt[emperature][=n]
dro[hms][=n]
drr[atio][=n]
dn[umber][=n]
dt[ype][=d[ata]/g[raph]]
dfp[arameter](d,f)[=n]
dfi[ndex](d,i)[=n]
grhm[ode](d)[=a[uto]/ s[et]]
grlm[ode](d)[=a[uto]/s[et]]
grhv[alue](d)[=n]
grlv[alue](d)[=n]
106
Description
read or set three-wire lead offset for probe
read or set ITS-90 low range selection
read or set ITS-90 high range selection
read or set ITS-90 R(TPW) probe constant
read or set ITS-90 a probe constants for range
read or set ITS-90 b probe constants for range
read or set ITS-90 c or ci probe constants for range
read or set ITS-90 d constant
read or set IPTS-68 R(0) constant
read or set IPTS-68 alpha constant
read or set IPTS-68 delta constant
read or set IPTS-68 a constant for range
read or set IPTS-68 b constant for range
read or set IPTS-68 c constant for range
read or set IPTS-68 d constant for range
read or select scale conversion, ITS-90 or IPTS-68
read or set Callendar Van-Dusen R(0) constant
read or set Callendar Van-Dusen alpha constant
read or set Callendar Van-Dusen delta constant
read or set Callendar Van-Dusen beta constant
read or set polynomial coefficients
read or set thermistor coefficients ai [T(R)]
read or set thermistor coefficients bi [R(T)]
read or set digits of resolution for temperature (C, K, F)
read or set digits of resolution for resistance (W)
read or set digits of resolution for ratio (r, W)
read or set display number
read or set display type
read or set display field parameter (see Section 5.6.3)
read or set display field memory/history index
read or set mode of high limit for graph display
read or set mode of low limit for graph display
read or set value of high limit for graph display
read or set value of low limit for graph display
Example
po(16)=0.2857
rl90(16)=4
rh90(16)=7
r90(16)=25.48635
a90(16,7)=-3.4837E-4
b90(16,7)=1.5385E-5
c90(16,7)=1.3820E-6
d90(16)=4.876534E-9
r68(16)=100.037465
al68(16)=0.00392583
de68(16)=1.495873
a68(16,4)=-3.8765E-6
b68(16,3)=8.2094E-8
c68(16,4)=1.4930E-7
d68(16,2)=-3.3654E-6
scale(16)=68
rcvd(16)=100.0
alcvd(16)=0.00385
decvd(16)=1.507
becvd(16)=0.111
ap(16,0)=31.9438576
at(16,0)=0.00368732
bt(16,0)=-4.4728857
drt=4
dro=6
drr=8
dn=16
dt=g
dfp(16,5)=3
dfi(16,5)=100
grhm(16)=s
grlm(16)=s
grhv(16)=100
grlv(16)=0