When START is pressed the Super-Thermometer will begin measuring the tem-
perature. The temperature (in Kelvin) will appear in the window. You may
graph the measurements by pressing the GRAPH soft-key. After allowing ade-
quate time for the measurement to stabilize, press the STOP soft-key (or Enter)
to stop measuring. If the TIMED measurement mode was selected in the setup,
measuring will stop automatically and the value stored after the specified
length of time has elapsed.
If the thermometer detects that the noise of the measurements exceeds the filter
threshold, a warning message will appear. To fix this, return, go into the FIL-
TER parameter window in the SAMPLING MENU, and increase the filter
threshold value. Start over with the calibration.
When STOP is pressed the final value of the measurement will be shown. Press
the STORE soft-key (or Enter) to save the value. The temperature measurement
will be stored with the calibration data for the given probe or for all probes de-
pending on the TEMP setting in the setup. If the SPRT is on INPUT 2 and the