DFG35 Digital Force Gauges
CALIBRATION
COMPRESSION
Attach necessary
weight fixtures,
then press ENTER.
Attach weight fixtures (brackets, hooks, etc), as required. Do not yet attach any weights or apply
any calibration loads. Then press ENTER.
7. The display will appear as follows:
CALIBRATION
COMPRESSION
Optionally exercise
sensor, then press
ENTER.
Optionally exercise the load cell shaft several times (at full scale, if possible), then press ENTER.
8. The display will appear as follows:
CALIBRATION
COMPRESSION
Gain adjust
Apply full scale load
10.00 lbF +/-20%,
then press ENTER.
Apply a weight equal to the full scale of the instrument, and then press ENTER.
9. After displaying "PLEASE WAIT..." the display will appear as follows:
CALIBRATION
COMPRESSION
Ensure no load,
then press ZERO.
Remove the load applied in Step 8, leave the fixtures in place, then press ZERO.
10. The display will appear as follows:
CALIBRATION
COMPRESSION
Apply load
1 OF 5
Enter load:
2.00 lbF
Press ENTER.