Instrument Settings

Use the Instrument Settings menu to change global instrument setting
parameters. Push  (Instrument Settings) to view the Instrument Settings
submenu. See Table 22.
Menu Parameter
Line Frequency
Ext. Ref. Clk
(External Reference
Clock)
Table 22. Instrument Settings Parameters
The Product automatically detects the mains frequency but a specific line
frequency can be set here in non-volatile memory. To change the Line
Frequency, use the navigation keys to move the cursor to highlight this field
and then push  (Auto) or  (50 Hz)  (60 Hz). The Auto setting
measures the line frequency at the time the Auto settings is made and at
Product power-on. It does not continually monitor the line frequency nor does it
use an off-nominal value.
When on, this enables the rear-panel FREQ REF IN input to use an external
1 MHz or 10 MHz clock. To change the External Reference Clock, use the
navigation keys to move the cursor to highlight this field and then push 
(OFF) or  (1MHz) or  (10MHz).
Reference Multimeter and 8 ½ Digit Multimeter
Change the Parameter
Instrument Setup
99