Endress+Hauser Proline Promass E 200 Technical Information

Coriolis flowmeter
Hide thumbs Also See for Proline Promass E 200:
Table of Contents

Advertisement

TI01009D/06/EN/17.14
71237197
The genuine loop-powered flowmeter for minimized cost of ownership
Application
• Measuring principle operates independently of physical fluid
properties such as viscosity or density
• Highly accurate measurement of liquids and gases for a wide
range of standard applications
Device properties
• Compact dual-tube system
• Medium temperature up to +140 °C (+284 °F)
• Process pressure up to 100 bar (1 450 psi)
• Loop-powered technology
• Robust two-chamber housing
• Plant safety: worldwide approvals (SIL, Haz. area)
Products
Technical Information
Proline Promass E 200
Coriolis flowmeter
Solutions
Your benefits
• Cost-effective – multi-purpose device; an alternative to
conventional volumetric flowmeters
• Fewer process measuring points – multivariable
measurement (flow, density, temperature)
• Space-saving installation – no in/outlet run needs
• Convenient device wiring – separate connection
compartment
• Safe operation – no need to open the device due to display
with touch control, background lighting
• Integrated verification – Heartbeat Technology™
Services

Advertisement

Table of Contents
loading

Summary of Contents for Endress+Hauser Proline Promass E 200

  • Page 1 Products Solutions Services TI01009D/06/EN/17.14 71237197 Technical Information Proline Promass E 200 Coriolis flowmeter The genuine loop-powered flowmeter for minimized cost of ownership Application Your benefits • Measuring principle operates independently of physical fluid • Cost-effective – multi-purpose device; an alternative to...
  • Page 2: Table Of Contents

    Proline Promass E 200 Table of contents Document information ..... . . 3 Vibration resistance ......31 Interior cleaning .
  • Page 3: Document Information

    Proline Promass E 200 Document information Symbols used Electrical symbols Symbol Meaning Direct current A terminal to which DC voltage is applied or through which direct current flows. A0011197 Alternating current A terminal to which alternating voltage is applied or through which alternating current flows.
  • Page 4: Function And System Design

    Proline Promass E 200 Symbol Meaning Flow direction A0013441 Hazardous area Indicates a hazardous area. A0011187 Safe area (non-hazardous area) Indicates a non-hazardous area. A0011188 Function and system design Measuring principle The measuring principle is based on the controlled generation of Coriolis forces. These forces are always present in a system when both translational and rotational movements are superimposed.
  • Page 5: Measuring System

    Proline Promass E 200 Temperature measurement The temperature of the measuring tube is determined in order to calculate the compensation factor due to temperature effects. This signal corresponds to the process temperature and is also available as an output signal.
  • Page 6: Operable Flow Range

    To increase the accuracy of certain measured variables or to calculate the corrected volume flow for gases, the automation system can continuously write the operating pressure to the measuring device. Endress+Hauser recommends the use of a pressure transmitter for absolute pressure, e.g. Cerabar M or Cerabar S.
  • Page 7: Output

    Proline Promass E 200 Fieldbuses The measured values can be written from the automation system to the measuring via: PROFIBUS-PA Output Output signal Current output Current output 1 4-20 mA HART (passive) Current output 2 4-20 mA (passive) Resolution <1 µA Damping Adjustable: 0.0 to 999.9 s...
  • Page 8: Signal On Alarm

    Proline Promass E 200 Number of switching Unlimited cycles Assignable functions • Off • On • Diagnostic behavior • Limit value – Mass flow – Volume flow – Corrected volume flow – Density – Reference density – Temperature – Totalizer 1-3 •...
  • Page 9: Load

    Proline Promass E 200 PROFIBUS PA Status and alarm Diagnostics in accordance with PROFIBUS PA Profile 3.02 messages Error current FDE (Fault 0 mA Disconnection Electronic) Local display Plain text display With information on cause and remedial measures Backlight Additionally for device version with SD03 local display: red lighting indicates a device error.
  • Page 10: Ex Connection Data

    Proline Promass E 200 Sample calculation Supply voltage of the supply unit: U = 19 V Maximum load: R ≤ (19 V - 13.5 V) :0.022 A = 250 Ω Ex connection data Safety-related values Ex d type of protection Order code for "Output"...
  • Page 11 Proline Promass E 200 Intrinsically safe values Type of protection Ex ia Order code for "Output" Output type Intrinsically safe values Option A 4-20mA HART = DC 30 V = 300 mA = 1 W = 0 μH = 5 nF...
  • Page 12: Low Flow Cut Off

    Proline Promass E 200 Order code for "Output" Output type Intrinsically safe values Option G PROFIBUS PA STANDARD FISCO = 32 V = 17.5 V = 300 mA = n.a. = n.a. = n.a. = 10 µH = 10 µH...
  • Page 13 Proline Promass E 200 HART load • Min. 250 Ω • Max. 500 Ω Dynamic variables The measured variables can be freely assigned to the dynamic variables. Measured variables for PV (primary dynamic variable) • Mass flow • Volume flow •...
  • Page 14: Power Supply

    Proline Promass E 200 Supported functions • Identification & Maintenance Simplest device identification on the part of the control system and nameplate • PROFIBUS upload/download Reading and writing parameters is up to ten times faster with PROFIBUS upload/download • Condensed status...
  • Page 15: Supply Voltage

    For device version with SD03 local display: The terminal voltage must be increased by DC 0.5 V if backlighting is used. For information about the load see (→  9) Various power supply units can be ordered from Endress+Hauser: see "Accessories" section (→  60) For information on the Ex connection values (→  10)
  • Page 16: Electrical Connection

    Proline Promass E 200 Electrical connection Connecting the transmitter A0015510 Cable entry for output 1 Cable entry for output 2 Connection examples Current output 4-20 mA HART 4...20 mA A0015511  1 Connection example for 4-20 mA HART current output (passive) Automation system with current input (e.g.
  • Page 17 Proline Promass E 200 Pulse/frequency output – – 12345 A0016801  3 Connection example for pulse/frequency output (passive) Automation system with pulse/frequency input (e.g. PLC) Power supply Transmitter: observe input values (→  7) Switch output A0016802  4 Connection example for switch output (passive) Automation system with switch input (e.g.
  • Page 18 Proline Promass E 200 PROFIBUS-PA A0019004  5 Connection example for PROFIBUS-PA Control system (e.g. PLC) Segment coupler PROFIBUS DP/PA Cable shield T-box Measuring device Local grounding Bus terminator Potential matching line HART input 4...20 mA – – – –...
  • Page 19: Potential Equalization

    Proline Promass E 200 Potential equalization No special measures for potential equalization are required. For devices intended for use in hazardous locations, please observe the guidelines in the Ex documentation (XA). • For device version without integrated overvoltage protection: plug-in spring terminals for wire Terminals cross-sections 0.5 to 2.5 mm...
  • Page 20: Performance Characteristics

    Proline Promass E 200 Performance characteristics Reference operating • Error limits based on ISO 11631 conditions • Water with +15 to +45 °C (+59 to +113 °F) at2 to 6 bar (29 to 87 psi) • Specifications as per calibration protocol •...
  • Page 21: Repeatability

    Proline Promass E 200 US units 1:10 1:20 1:50 1:100 1:500 [inch] [lb/min] [lb/min] [lb/min] [lb/min] [lb/min] [lb/min] ³⁄₈ 73.5 7.35 3.675 1.47 0.735 0.147 ½ 23.8 11.9 4.76 2.38 13.2 1.32 1½ 1 650 82.5 16.5 2 570 128.5 51.4...
  • Page 22: Influence Of Medium Temperature

    Proline Promass E 200 Influence of medium Mass flow and volume flow When there is a difference between the temperature for zero point adjustment and the process temperature temperature, the typical measured error of the sensor is ±0.0002 % of the full scale value/°C (±0.0001 % of the full scale value/°F).
  • Page 23: Installation

    Proline Promass E 200 Calculation of the maximum repeatability as a function of the flow rate Flow rate Maximum repeatability in % o.r. ⋅ ⁄ ZeroPoint ⋅ ⋅ ± ½ BaseAccu ³ BaseAccu A0021343 A0021341 ⋅ ⁄ ZeroPoint ZeroPoint ⋅...
  • Page 24: Orientation

    Proline Promass E 200 A0015596  9 Installation in a down pipe (e.g. for batching applications) Supply tank Sensor Orifice plate, pipe restriction Valve Batching tank Ø orifice plate, pipe restriction [mm] [in] [mm] [in] ³⁄₈ 0.24 ½ 0.40 0.55 1½...
  • Page 25: Inlet And Outlet Runs

    Proline Promass E 200 If a sensor is installed horizontally with a curved measuring tube, match the position of the sensor to the fluid properties. A0014057  10 Orientation of sensor with curved measuring tube Avoid this orientation for fluids with entrained solids: Risk of solids accumulating.
  • Page 26 Proline Promass E 200 Avoid direct sunlight, particularly in warm climatic regions. Weather protection covers can be ordered from Endress+Hauser: see "Accessories" section (→  57) Temperature tables In the following tables, the following interdependencies between the maximum medium temperature for T1-T6 and the maximum ambient temperature T apply when operating the device in hazardous areas.
  • Page 27 Proline Promass E 200 US units Nominal diameter [in] [°F] [185 °F] [212 °F] [275 °F] [392 °F] [572 °F] [842 °F] 1) 2) ³⁄₈ to 2 3) 2) ³⁄₈ to 2 – ³⁄₈ to 2 – – = 104 °F for pulse/frequency/switch output P ≤...
  • Page 28 Proline Promass E 200 US units Nominal diameter [in] [°F] [185 °F] [212 °F] [275 °F] [392 °F] [572 °F] [842 °F] ³⁄₈ to 2 ³⁄₈ to 2 – ³⁄₈ to 2 – – The following applies for installations with overvoltage protection in conjunction with temperature class T5, T6: T - 3.6 °F...
  • Page 29 Proline Promass E 200 US units Nominal diameter [in] [°F] [185 °F] [212 °F] [275 °F] [392 °F] [572 °F] [842 °F] ³⁄₈ to 2 2) 3) ³⁄₈ to 2 – ³⁄₈ to 2 – – = 122 °F without pulse/frequency/switch output = 131 °F without pulse/frequency/switch output...
  • Page 30: Storage Temperature

    Proline Promass E 200 US units Nominal diameter [in] [°F] [185 °F] [212 °F] [275 °F] [392 °F] [572 °F] [842 °F] ³⁄₈ to 2 2) 3) ³⁄₈ to 2 – ³⁄₈ to 2 – – = 122 °F without pulse/frequency/switch output = 131 °F without pulse/frequency/switch output...
  • Page 31: Shock Resistance

    Proline Promass E 200 Sensor IP66/67, type 4X enclosure Device plug IP67, only in screwed situation Shock resistance As per IEC/EN 60068-2-31 Vibration resistance Acceleration up to 1 g, 10 to 150 Hz, based on IEC/EN 60068-2-6 Interior cleaning • SIP cleaning •...
  • Page 32 Proline Promass E 200 Flange connection according to ASME B16.5 [psi] [bar] 1400 1200 Class 600 1000 Class 300 Class 150 [°C] [°F] A0020973-EN  14 With flange material 1.4404 (F316/F316L) Flange connection as per JIS B2220 [psi] [bar] 1000 [°C]...
  • Page 33 Proline Promass E 200 VCO process connection [psi] [bar] 1400 1200 1000 [°C] [°F] A0020975-EN  16 With connection material 1.4404 (316/316L) Tri-Clamp The clamp connections are suitable up to a maximum pressure of 16 bar (232 psi). Please observe the operating limits of the clamp and seal used as they could be under 16 bar (232 psi).
  • Page 34: Secondary Containment Pressure Range

    Proline Promass E 200 SMS 1145 allows for applications up to 6 bar (87 psi) if suitable sealing materials are used. Please take this into account when selecting seals and counterparts, as these components can limit the pressure and temperature range.
  • Page 35: Rupture Disk

    • Via pipes carrying hot water or steam • Via heating jackets Heating jackets for the sensor can be ordered as accessories from Endress+Hauser (→  57). Vibrations The high oscillation frequency of the measuring tubes ensures that the correct operation of the measuring system is not influenced by plant vibrations.
  • Page 36: Mechanical Construction

    Proline Promass E 200 Mechanical construction Design, dimensions Compact version Order code for "Housing", options B "GT18 two-chamber, 316L", C "GT20 two-chamber aluminum coated" A0013552 Dimensions in SI units for version without overvoltage protection [mm] [mm] [mm] [mm] [mm] [mm]...
  • Page 37 Proline Promass E 200 Process connections in SI units Flange connections EN (DIN) +1,5 (+0.06) -2,0 (-0.08) A0015621  22 Engineering unit mm (in) Flange according to EN 1092-1 (DIN 2501 / DIN 2512N ) / PN 40: 1.4404 (F316/F316L) (order code for "Process connection", option D2S)
  • Page 38 Proline Promass E 200 Flange according to EN 1092-1 (DIN 2501 / DIN 2512N ) / PN 100: 1.4404 (F316/F316L) (order code for "Process connection", option D4S) Surface roughness (flange): EN 1092-1 Form B2 (DIN 2526 Form E), Ra 0.8 to 3.2 µm...
  • Page 39 Proline Promass E 200 Flange according to ASME B16.5 / Cl 300: 1.4404 (F316/F316L) (order code for "Process connection", option ABS) [mm] [mm] [mm] [mm] [mm] [mm] [mm] 155.4 114.3 4 × Ø22.3 20.6 40.9 165.1 127.0 8 × Ø19.0 22.3...
  • Page 40 Proline Promass E 200 Flange JIS B2220 / 20K: 1.4404 (F316/F316L) (order code for "Process connection", option NES) [mm] [mm] [mm] [mm] [mm] [mm] [mm] 4 × Ø19 8 × Ø19 DN 8 with DN 15 flanges as standard Flange JIS B2220 / 40K: 1.4404 (F316/F316L) (order code for "Process connection", option NGS)
  • Page 41 Proline Promass E 200 VCO connections: 1.4404 (316/316L) [mm] [in] [mm] [mm] AF 1 10.2 AF 1½ 15.7 8-VCO-4 (½"): (order code for "Process connection", option CVS) 12-VCO-4 (¾"): (order code for "Process connection", option CWS) Tri-Clamp +1,5 (+0.06) -2,0 (-0.08) A0015625 ...
  • Page 42 Proline Promass E 200 DIN 11851 (sanitary connection) +1,5 (+0.06) -2,0 (-0.08) A0015628  27 Engineering unit mm (in) Sanitary connection DIN 11851: 1.4404 (316/316L) (order code for "Process connection", option FMW) [mm] [in] [mm] [mm] Rd 34 × ¹⁄₈...
  • Page 43 Proline Promass E 200 Threaded hygienic connection DIN 11864-1 Form A: 1.4404 (316/316L) (order code for "Process connection", option FLW) [mm] [in] [mm] [mm] Rd 78 × ¹⁄₆ 3A version also available (Ra ≤ 0.8 µm) DIN 11864-2 Form A (flange with groove) +1,5 (+0.06)
  • Page 44 Proline Promass E 200 ISO 2853 (threaded hygienic connection) +1,5 (+0.06) -2,0 (-0.08) A0015623  30 Engineering unit mm (in) Threaded hygienic connection ISO 2853: 1.4404 (316/316L) (order code for "Process connection", option JSF) [mm] [mm] [mm] [mm] 37.13 22.6 37.13...
  • Page 45 Proline Promass E 200 Threaded hygienic connection SMS 1145: 1.4404 (316/316L) (order code for "Process connection", option FSW) [mm] [in] [mm] [mm] Rd 40 × ¹⁄₆ 22.5 Rd 60 × ¹⁄₆ 35.5 Rd 70 × ¹⁄₆ 48.5 3A version also available (Ra ≤ 0.8 µm)
  • Page 46 Proline Promass E 200 Process connections in US units Flange connections ASME B16.5 +1,5 (+0.06) -2,0 (-0.08) A0015621  32 Engineering unit mm (in) Flange according to ASME B16.5 / Cl 150: 1.4404 (F316/F316L) (order code for "Process connection", option AAS)
  • Page 47 Proline Promass E 200 Flange according to ASME B16.5 / Cl 600: 1.4404 (F316/F316L) (order code for "Process connection", option ACS) [in] [in] [in] [in] [in] [in] [in] 1½ 6.12 4.50 4 × Ø0.88 1.13 1.50 19.53 6.50 5.00 8 × Ø0.75 1.25...
  • Page 48 Proline Promass E 200 1", 1½", 2" Tri-Clamp for pipe size: 1.4404 (316/316L) (order code for "Process connection", option FTS) Clamp [in] [in] [in] [in] [in] ³⁄₈ 1.98 0.87 9.02 ½ 1.98 0.87 10.75 1.98 0.87 12.76 1½ 1½ 1.98 1.37...
  • Page 49: Weight

    Proline Promass E 200 Weight Compact version Weight in SI units All values (weight) refer to devices with EN/DIN PN 40 flanges. Weight information in [kg]. Weight [kg] [mm] Order code for "Housing", option C Order code for "Housing", option B Aluminum coated 1.4404 (316L)
  • Page 50 Proline Promass E 200 Order code for "Housing", option B "GT18 two-chamber, 316L" Cable entry/cable gland Type of protection Material Cable gland M20 × 1.5 • Non-Ex Stainless steel 1.4404 • Ex ia • Ex ic • Ex nA • Ex tb...
  • Page 51: Process Connections

    Proline Promass E 200 Seals Welded process connections without internal seals Accessories Weather protection cover Stainless steel 1.4301 Process connections • Flanges: – EN 1092-1 (DIN 2501) – Namur lengths in accordance with NE 132 – ASME B16.5 – JIS B2220 •...
  • Page 52: Local Operation

    Proline Promass E 200 Local operation Via display module Order code for "Display; Operation", option C "SD02" Order code for "Display; Operation", option E "SD03" A0015546 A0015544 Operation with pushbuttons Operation with touch control Display elements • 4-line display • With order code for "Display; operation", option E: White background lighting;...
  • Page 53: Remote Operation

    VIATOR Bluetooth modem with connecting cable Transmitter Via service interface (CDI) A0014019 Service interface (CDI = Endress+Hauser Common Data Interface) of the measuring device Commubox FXA291 Computer with "FieldCare" operating tool with COM DTM "CDI Communication FXA291" Via PROFIBUS PA network This communication interface is present in the following device version: Order code for "Output", option G: PROFIBUS PA...
  • Page 54: Certificates And Approvals

    Directives. These are listed in the corresponding EC Declaration of Conformity along with the standards applied. Endress+Hauser confirms successful testing of the device by affixing to it the CE mark. C-Tick symbol The measuring system meets the EMC requirements of the "Australian Communications and Media Authority (ACMA)".
  • Page 55: Hygienic Compatibility

    Proline Promass E 200 Ex ia Category (ATEX) Type of protection II1/2G Ex ia IIC T6…T1 Ga/Gb II2G Ex ia IIC T6…T1 Gb II1/2G, II2D Ex ia IIC T6…T1 Ga/Gb Ex tb IIIC Txxx Db Ex nA Category (ATEX) Type of protection...
  • Page 56: Other Standards And Guidelines

    Proline Promass E 200 • With the PED/G1/x (x = category) marking on the sensor nameplate, Endress+Hauser confirms compliance with the "Essential Safety Requirements" specified in Annex I of the Pressure Equipment Directive 97/23/EC. • Devices bearing this marking (PED) are suitable for the following types of medium: –...
  • Page 57: Application Packages

    The application packages can be ordered from Endress+Hauser either directly with the device or subsequently. Detailed information on the order code in question is available from your local Endress +Hauser sales center or on the product page of the Endress+Hauser website: www.endress.com.
  • Page 58: Communication-Specific Accessories

    Is used to stabilize the temperature of the fluids in the sensor. Water, water vapor and other non-corrosive liquids are permitted for use as fluids. If using oil as a heating medium, please consult with Endress+Hauser. Heating jackets cannot be used with sensors fitted with a rupture disk.
  • Page 59: Service-Specific Accessories

    The application already contains the data of your Endress+Hauser device. Endress +Hauser also takes care of maintaining and updating the data records.
  • Page 60: System Components

    • The W@M Device Viewer : Enter the serial number from the nameplate (www.endress.com/deviceviewer) • The Endress+Hauser Operations App: Enter the serial number from the nameplate or scan the 2-D matrix code (QR code) on the nameplate. Standard documentation...
  • Page 61: Registered Trademarks

    Registered trademark of the PROFIBUS User Organization, Karlsruhe, Germany TRI-CLAMP ® Registered trademark of Ladish & Co., Inc., Kenosha, USA Applicator ® , FieldCare ® , Field Xpert , HistoROM ® , Heartbeat Technology Registered or registration-pending trademarks of the Endress+Hauser Group Endress+Hauser...
  • Page 64 www.addresses.endress.com...