Operating and Service Instruction

Operator's Check
C A U T I O N
Agilent 8710x Operating and Service Manual
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com
The operator's check is supplied to allow the operator to make a
quick check on the coaxial multiport switches prior to use or if a
failure is suspected.
ESD exceeding the level specified in
power applied is greater than the maximum specified as in
Specifications"
may cause permanent damage to the device.
Description
The coaxial multiport switch is connected to a network analyzer
configured for the s- parameter measurement. The network analyzer
may be set to sweep over the whole or selected frequency range of
the switch to be verified. The s- parameters measurement is the
best way to determine if the switch is working properly.
NETWORK ANALYZER
Port 1
Figure 8
Connection to Perform Quick Check
Installation and Verification
"Environmental Specifications"
POWER SUPPLY
Port 2
Agilent Coaxial
Multiport Switch
4
or RF
"General
27