Alarm   T Ype   ( Prog   >   A Lm.1,   A Lm.2   >   T Ype) - Omega PLATINUM Series User Manual

Temperature, process and strain meters
Hide thumbs Also See for PLATINUM Series:
Table of Contents

Advertisement

PLATINUM   S eries   M eters   U ser's   G uide  
 
 
7.3
Alarm   M ode   C onfiguration   ( PRoG   >   A LM.1,   A LM.2)  
Select   A larm   C onfiguration   1   ( ALM.1)   o r   A larm   C onfiguration   2   ( ALM.2)   i n   o rder   t o   s et   u p,  
 
change,   e nable,   o r   d isable   A larms.   E ither   o r   b oth   A larms   c an   b e   a ssigned   t o   t rigger   d isplay  
color   c hanges,   a nnunciators,   a nd   /   o r   o utputs.   E ither   o r   b oth   A larm   c onfigurations   c an   b e  
assigned   t o   m ultiple   o utputs.     T he   A LM.1   a nd   A LM.2   c onfiguration   m enus   h ave   a ll   o f   t he  
same   s ettings   a nd   f unction   i n   t he   s ame   m anner.  
Navigate   t o   t he   A larm   s etting   y ou   w ant   t o   c hange.   S ettings   i nclude   t he   f ollowing:  
 
tyPE  
Ab.dV   –   Alarm   r eferences   v alues   ( ALR.H   a nd   A LR.L)   o r   d eviation   f rom   S P1   o r   S P2  
ALR.H   –   Alarm   h igh   p arameter,   u sed   f or   A larm   t rigger   c alculations  
ALR.L   –   Alarm   l ow   p arameter,   u sed   f or   A larm   t rigger   c alculations  
A.CLR   –   Alarm   c olor   i ndication  
HI.HI  
LtCH  
CtCL  
A.P.oN   –   Alarm   p ower-­‐on   b ehavior  
dE.oN   –   Time   d elay   f or   A larm   t rigger   u nless   t he   c ondition   p ersists,   d efault   =   1 .0   s  
dE.oF   –   Time   d elay   f or   c ancelling   A larms   a fter   b eing   t riggered;   p revents   A larm  
"chatter,"   d efault   =   0 .0   s  
Select   t he   i ndicated   s etting.  
 
 
7.3.1
Alarm   T ype   ( PRoG   >   A LM.1,   A LM.2   >   t yPE)  
Select   t he   A larm   T ype   ( tyPE)   p arameter.     T his   p arameter   w ill   c ontrol   t he   b asic   b ehavior   o f   t he  
 
selected   a larm.  
Navigate   t o   t he   d esired   s etting.   S ettings   i nclude   t he   f ollowing:  
 
oFF  
–   Alarm   i s   o ff   ( factory   d efault)  
AboV   –   Alarm   i s   t riggered   w hen   t he   p rocess   v alue   e xceeds   A LR.H   ( Absolute   M ode)  
or   t he   s pecified   S etpoint   p lus   A LR.H   ( Deviation   M ode)  
bELo   –   Alarm   i s   t riggered   w hen   t he   p rocess   v alue   i s   l ess   t han   A LR.L   ( Absolute  
Mode)   o r   t he   s pecified   S etpoint   m inus   A LR.L   ( Deviation   M ode)  
HI.Lo.   –   Alarm   i s   t riggered   w hen   t he   p rocess   v alue   i s   o utside   t he   A LR.L–ALR.H   r ange  
(Absolute   M ode)   o r   t he   r ange   d efined   b y   t he   b and   a round   t he   s pecified   S etpoint   a s  
determined   b y   A LR.L   a nd   A LR.H   ( Deviation   M ode)    
bANd   –   Alarm   i s   t riggered   w hen   t he   p rocess   v alue   i s   w ithin   t he   A LR.L–ALR.H   r ange  
(Absolute   M ode)   o r   w ithin   t he   b and   a round   t he   s pecified   S etpoint   a s   d etermined   b y  
ALR.L   a nd   A LR.H   ( Deviation   M ode)  
Note:   T able   5 .1   c ompares   t he   A larm   r ange   o ptions,   a nd   F igure   5 .1   r epresents   t he   A larm   r ange  
options   g raphically.  
Select   t he   i ndicated   s etting.  
 
 
–   Alarm   t ype   a bsolute   o r   d eviation  
–   High   H igh   /   L ow   L ow   o ffset   v alue  
–   Alarm   l atching  
–   Alarm   a ction   ( normally   o pen   o r   n ormally   c losed)  
 
Reference   S ection:   P rogramming   M ode   ( PRoG)  
44  
 

Hide quick links:

Advertisement

Table of Contents
loading