Thermistor   I Nput   T Ype   C Onfiguration   ( Init   >   I Npt   >   T Hrm); Resistance   T Emperature   D Etector   ( Rtd)   I Nput   T Ype   ( Init   >   I Npt   >   R Td)   2 - Omega PLATINUM Series User Manual

Temperature, process and strain meters
Hide thumbs Also See for PLATINUM Series:
Table of Contents

Advertisement

PLATINUM   S eries   M eters   U ser's   G uide  
 
 
6.1.2
Resistance   T emperature   D etector   ( RTD)   I nput   T ype   ( INIt   >   I NPt  
>   R td)  
Select   R td   a s   t he   i nput   t ype.   F actory   d efault   c onfiguration   s ettings   a re   t hree-­‐wire,   1 00   Ω ,  
 
using   t he   E uropean   s tandard   3 85   c urve.   N ote   t hat   t he   3 92   a nd   3 916   c urves   a re   o nly   a vailable  
for   1 00   Ω   R TDs.   I f   R td   i s   s elected   a nd   a   s pecific   c onfiguration   i s   n ot   c hanged,   t he   l ast   s aved  
configuration   w ill   b e   u sed.  
Navigate   t o   t he   d esired   c onfiguration   p arameter:  
 
N.wIR   –   Firmware   s election   o f   t he   n umber   o f   w ires   f or   R TD   c onnection   ( no   j umpers  
needed)  
A.CRV   –   Calibration   c urve   c overing   b oth   t he   i nternational   s tandard   a nd   t he  
resistance   o f   t he   R TD  
Select   t he   o ption.  
 
Navigate   t o   t he   d esired   s etting.   S ettings   i nclude   t he   f ollowing:  
 
Select   t he   i ndicated   o ption.  
 
Navigate   t o   t he   d esired   s etting.   S ettings   i nclude   t he   f ollowing:  
 
Select   t he   i ndicated   o ption.  
 
6.1.3
Thermistor   I nput   T ype   C onfiguration   ( INIt   >   I NPt   >   t HRM)  
Select   T hermistor   ( tHRM)   a s   t he   i nput   t ype.   T his   s ets   u p   t he   u nit   f or   t hermistor-­‐based  
 
temperature   m easurement   a nd   t hen   t he   s pecific   t hermistor   t ype   c an   b e   s pecified.   I f   n o  
thermistor   t ype   i s   s pecified,   t he   l ast   s elected   t ype   i s   u sed.  
Navigate   t o   t he   c orrect   s etting.   S ettings   i nclude   t he   f ollowing:  
 
2.25k  –   2,250   Ω   t hermistor   ( factory   d efault)  
5k  
–   5,000   Ω   t hermistor  
10k   –   10,000   Ω   t hermistor  
Select   t he   i ndicated   o ption.  
 
 
6.1.2.1 Number   o f   R TD   W ires   ( INIt   >   I NPt   >   R td   >  
N.wIR)  
3   w I   –   Three-­‐wire   R TD   ( factory   d efault)  
4   w I   –   Four-­‐wire   R TD  
2   w I   –   Two-­‐wire   R TD  
6.1.2.2 Calibration   C urve   ( INIt   >   I NPt   >   R td   >   A .CRV)  
385.1   –   European   a nd   m ost   c ommon   s tandard   a t   t he   c onventional  
resistance   o f   1 00   Ω   ( factory   d efault)  
385.5   –   European   c urve   f or   5 00   Ω  
385.t   –   European   c urve   f or   1 000   Ω  
392   –   Old   U S   s tandard   ( rarely   u sed),   a t   1 00   Ω   o nly  
3916   –   Japanese   s tandard,   a t   1 00   Ω   o nly  
Reference   S ection:   I nitialization   M ode   ( INIt)  