Vibration Resistance - Siemens 7MF03 Series Compact Operating Instructions

Pressure transmitter with 4 to 20 ma/hart
Table of Contents

Advertisement

Operating conditions for level
Electromagnetic compatibili‐
ty
Interference emission and
interference immunity
Process medium conditions
● Process temperature
Measuring cell with sili‐
cone oil filling
8.5

Vibration resistance

General operating conditions
Vibrations (sine)
IEC 60068-2-6
Continuous shocks (half-sine)
IEC 60068-2-27
Continuous shocks (half-sine)
IEC 60068-2-27
1) Without mounting bracket
2) With mounting bracket
SITRANS P320/P420 with 4 to 20 mA/HART
Compact Operating Instructions, 09/2018, A5E38874562-AB
Enclosure with mounted HAN con‐
nector
In accordance with EN 61326 and
NAMUR NE 21
● Plus side: See mounting flange
● Minus side: -40 ... +100 °C
(-40 ... +212 °F)
Gauge pressure series
Aluminum and stainless steel enclosure
IP65
Differential pressure series
2)
Aluminum and stainless steel enclosure
2 ... 9 Hz at 0.3 mm
9 ... 200 Hz at 5 m/s²
1 octave/min
5 cycles/axis
70 m/s²
30 ms
6 shocks/axis
250 m/s²
6 ms
1000 shocks/axis
Technical data
8.5 Vibration resistance
1)