Invacare® Platinum™Series

6.2 Cleaning the Cabinet Filter

CAUTION!
Risk of Damage
To avoid damage from clogging:
– DO NOT operate the concentrator without the filter
installed.
There are two cabinet filters located on each side of the
cabinet.
24
1.
Remove the filter A and clean as needed.
Environmental conditions that may require more
frequent inspection and cleaning of the filter include, but
are not limited to: high dust, air pollutants, etc.
2.
Clean the cabinet filter with a vacuum cleaner or wash with a
mild liquid dish detergent (such as Dawn™) and water. Rinse
thoroughly.
3.
Thoroughly dry the filter and inspect for fraying, crumbling, tears
and holes. Replace filter if any damage is found.
4.
Reinstall the cabinet filter.
1195661-A