Invacare Platinum IRC9LXO2AWQ User Manual page 5

Homefill system compatible oxygen concentrators with senso2
Hide thumbs Also See for Platinum IRC9LXO2AWQ:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

Call supplier
**This product complies with Directive
93/42/EEC concerning medical devices.
The launch date of this product is stated in the
CE declaration of conformity.
For more details about CE marks, refer to Typical Product
Parameters.
Symbols on Product
Unit running
Unit not running
O
Indicators
2
Symbol
O
Purity
2
O
SYSTEM OKAY
2
O
over 85%
2
O
between 73% to
2
85%
SYSTEM FAILURE
O
Below 73%
2
1195661-A
Indicator Lights (LED)
GREEN Indicator Light
YELLOW Indicator Light
A. YELLOW Solid
B. YELLOW Flashing Sensor
Failure. Call a qualified
technician
RED Indicator Light
Refer to Troubleshooting.
LX Indicators
Symbol
O
Purity
2
I/O
SYSTEM OKAY
SYSTEM FAILURE
General
Indicator Lights (LED)
GREEN Indicator Light
RED Indicator Light
Continuous Audible Alarm
Sieve — GARD™
Compressor Shutdown
Call a qualified technician.