Your oxygen concentrator will perform best when operated under
the conditions outlined in the table below.
Usage in environments other than those described may result in the
need for increased equipment maintenance. The air intake of the
unit should be located in a well ventilated area to avoid airborne
pollutants and/or fumes.
Recommended Guidelines for Optimum Performance
Temperature:
Electrical:
Placement:
Tubing and Cannula:
Environment:
Relative Humidity:
1195661-A
10°C - 35°C (50°F - 95°F)
No extension cords.
No closer than 30,5 cm (12
in) from the wall, furniture,
draperies, or similar surfaces.
IRC5LXAW, IRC5LXO2AW,
IRC5LXO2AWQ
2,1 m (7 ft) cannula with a
maximum 15,2 m (50 ft) of
Crush-Proof Tubing (DO NOT
pinch.)
IRC9LXO2AWQ
Recommended use up to 15.2
m (50 ft) high flow tubing with
high flow cannula at all flow
rates.
Smoke and soot-free. No
confined spaces
(Example: No closets).
20 to 60%
Time of Operation:
Flow Rate
Minimum Operating Time:
Ensure that your concentrator is at least 30,5 cm (12 in)
away from walls, draperies or furniture to assure sufficient
air flow. Avoid deep pile carpets and heaters, radiators or
hot air registers.
Usage
Up to 24 hours per day.
IRC5LXAW, IRC5LXO2AW,
IRC5LXO2AWQ
0.5 L/min. to 5 L/min. Flow
rates less than 1 L/min., are not
recommended.
IRC9LXO2AWQ
1 L/min. to 9L/min. Flow rates
less than 1 L/min. are not
recommended.
30 Minutes