Quart Healthcare Topaz 8 Plus User Manual page 19

Alternating + micro low air loss pressure relief system
Hide thumbs Also See for Topaz 8 Plus:
Table of Contents

Advertisement

EMC INFORMATION (120V)
Guidance and manufacturer's declaration-electromagnetic emissions
The device(s) is intended for use in the electromagnetic environment specified
below. The customer or the user of the device(s) should assure that it is used in
such environment.
Emission test
environment.
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations
/flicker emissions
IEC 61000-3-3
Compliance
Electromagnetic environment-guidance
Group 1
The device(s) uses RF energy only for its
internal function. Therefore, its RF emissions
are very low and are not likely to cause any
interference in nearby electronic equipment.
Class B
The device(s) is suitable for use in all
establishments, including domestic
Class A
establishments and those directly connected
to the public low-voltage power supply
Compliance
network that supplies buildings used for
domestic purposes.