Performance Specifications

Flow Rate Measurement
Method ........................... Flow is calculated by
Range ............................ 0.1 to 1500 ml/h (2500 ml/h
Accuracy ........................ 1 % of reading ±1 LSD for
Max Test Duration ........ 100 hours
measuring volume over time
is shown)
flows of 16 to 200 ml/h for
volumes over 20 ml,
otherwise 2 % of reading ±1
LSD for volumes over 10 ml
under laboratory conditions.
Degassed water at 15 to
30 °C (59 to °86 F) is
recommended for long tests.
Accuracy for >1500 ml/hr not
specified.
Volume Measurement
Method ........................ Volume is measured directly
Range .......................... 0.06 ml to 9999 ml
Accuracy ..................... 1 % of reading ±1 LSD for
Max Test Duration ....... 100 hours
PCA Bolus/Dual Flow Measurement
Method ........................ See Volume measurement
Min Bolus Volume ....... 0.5 ml
Resolution ................... 60 μl increments
Max Test Duration ....... 100 hours
Infusion Device Analyzer
Specifications
by the measuring module in
minimum sample sizes of
60 μl
flow rates of 16 to 200 ml/h
for volumes over 20 ml.
otherwise 2 % of reading
±1 LSD for volumes over
10 ml under laboratory
conditions.
above
13